One,Hot Deformation Vickers Temperature TesterTechnical Specifications
1. Temperature control and measurement range: Room temperature – 300℃
2. Heating Rate: 120℃/h (12±1℃/6min) and 50℃/h (5±0.5℃/6min)
3. Temperature Resolution: 0.1℃
4. Large temperature error: 1℃
5. Deformation measurement range: 0-3mm, Large error: 0.01mm, Resolution: 0.001mm
6. Hot Deformation Test Span: 60mm - 180mm
7. Heating Medium: Methyl silicone oil (lower viscosity preferred) or transformer oil (for use up to 200℃)
8. Heating Power: 220V ±10% 20A 50Hz-60Hz
9. Main Unit Appearance: Vertical (with automatic rise and fall of the sample holder)
Two,Hot Deformation Vickers Temperature TesterPerformance Features
a) Real-time display of test temperature and temperature-deformation curve
b) During the test process, the required load mass for the test can be automatically calculated by inputting the sample's dimensions.
c) Online temperature and deformation amount can be set arbitrarily within the operating range.
d) Errors are automatically corrected through software
e)试验曲线 colors can be set arbitrarily, allowing for real-time display and hiding, as well as adjustable line widths.
(f) The alarm feature activates and heating automatically stops upon completion of the test or when the upper temperature limit is reached.
g) After completion of the test, print the test curve and test report. Query the historical records of past tests and be able to redraw the test curves for each past test.
h) Users can set a three-tier password protection for their software usage.