Rico Micro FT-341 Universal Dual Power Four-Probe Tester
Application InstructionsW:
Overcoat Film;Conductive polymeric film, high and low temperature electrothermal film;Insulating and Conductive Window Films - Conductive(Shield)Fabric, decorative film, decorative paper;Metalized labels, alloy foil;Melting, sintering, sputtering, coating, film coating, resistive, capacitive touch screen film;Electrode coatings, resistance measurement for other semiconductor materials, and thin film materials
Silicon ingots, crystal resistivity, diffusion layer, epitaxial layerITOConductive foil, conductive rubber, and other material resistance blocks Semiconductor Materials/Wafers, solar cells, electronic components, conductive filmsITOConductive film glass, metal films, conductive paint films, evaporated aluminum films.PCBCopper foil filmEMICoating material's thin layer resistivity and resistivity Conductive paint, conductive paste, conductive plastic, conductive rubber, conductive film, metallic film EMI Protective materials, conductive fibers, conductive ceramics, etc.Offer a choice of Chinese or English operating interface languages.

II.Description:
Rike Micro FT-341 General Dual Power Four Probes Tester
Utilizing a four-probe combination dual-electric measurement method, with a liquid crystal display, automatic measurement, automatic range, and automatic coefficient compensation..High-integration circuit system, constant current output; Optional:PCSoftware for data management and processing.
The dual-electrode digital four-probe tester utilizes dual-position measurement with linear or square four probes. The design of this instrument refers to the national standard physical testing methods for single crystal silicon.Please provide the Chinese content to be translated. A.S.T.M Standard. Employing current and voltage probe transformations, conduct two electrical measurements, perform double electrical measurement analysis on the data, and address the influence of factors such as sample geometric dimensions, boundary effects, and uneven probe spacing and mechanical drift on the measurement results..
Three.Parameters and specifications
Rike Micro FT-341 Universal Dual Power Four-Probe Tester
Model and Specifications FT-341
1.Square Resistor 10^-5~2×10^5Ω/□
2.Resistivity 10^-6~2×10^6Ω-cm
3.Test Current 0.1μA.μA.0μA,100µA,1mA,10mA,100mA
4.Current Accuracy ±0.1%
5.Resistance Accuracy ≤0.3%
6.Display Reading LCD Screen Display: Resistance, Resistivity, Square Resistance, Temperature, Unit Conversion, Temperature Coefficient, Current, Voltage, Probe Shape, Probe Spacing, Thickness Conductivity
7.Testing Method Dual Power Measurement
8.Power Supply Input: AC 220V±10%.50Hz Power Consumption:<30W
9.Tolerance ≤3%(Standard Sample Result)
10.Select Features Select1.pcSoftware; Selection2.Square probe; Select and purchase3.Linear probe; Select4.Test Platform5.Standard Resistor.
11.Test Probe Probe Spacing Selection:1mm;2mm;3mmThree specifications; Tungsten carbide probe material selection;White steel needle; gold-plated phosphor bronze hemispherical needle
12.Standard Resistors (Select) Specification:1mΩ、10mΩ、100mΩ、1Ω、10Ω、1kΩ、10kΩ、100kΩ、1MΩ