Product Introduction
This product is a laboratory-standard fully automatic fermenter, capable of meeting general testing requirements. It is widely used and features comprehensive functions.
1.1 System Composition
Serial Number | System | Dimensions (L) | Tank Material | Spiral Jacket/Multi-Tube Material | Quantity | Note |
1. | Fermenter | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Function
Serial Number | System/Equipment | Can lidLifting & Lowering | Control Function | |||||||
Temperature | Ph | DO | RPM (Revolutions Per Minute) | Traffic | Pressure | Sterilization | Supplementary Material | |||
1. | Fermentation Vessel | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Self-controlled | Self-regulating | Automatic | Solenoid Pumps |
1.3. Control Parameters
Serial Number | Control Parameters | Technical Specifications |
1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode compatible accessories Measurement Accuracy: 0.3 grade (0.3% FS ± 0.1℃), Measurement Range: 0-150℃ Control Method: Digital Setting, Manual/Automatic Control Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps Content Control: Bidirectional Heating and Cooling Control |
2. | PH | Swiss Company PH Glass Gel Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Stainless Steel pH Electrode Sheath for Matching Measurement Range: 0-14PH, Accuracy: ±0.01PH Four-digit display transmitter, powered by AC 220V, 4-20mA Original Swiss进口 electrode shielding wire equipped |
3. | DO | Swiss Company DO Membrane Electrode Electrodes can be sterilized in situ at high temperatures, sterilization temperature: 130°C. Matching stainless steel DO electrode sheath Measurement range: 0-100%, accuracy ±1% Supplied with a four-digit display transmitter, powered by AC 220V, 4-20mA. |
4. | RPM | Variable Frequency Infinite Speed Regulation System, power matched to the motor Digitalized setting of speed values Actuator: Variable Frequency Drive |
5. | Sterilization | Digitally set the sterilization temperature and time inside the vessel, with automatic control of the sterilization temperature and cooling process. Sterilization temperature range: 80℃-125℃. Sterilization Process: Steam is heated through the jacket to preheat the tank; once the preheating set temperature is reached, you can opt to set the steam to enter the tank directly. Achieve sterilization temperature inside the can; start the timer and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. |
6. | Pressure | Digitally set the tank pressure, with a 4-20mA signal for linear adjustment of the tank pressure. |
7. | Traffic (optional) | Flowmeter-assisted control of air flow into the tank |
8. | Defoamer | Control Method: Manual/Automatic Control Control Content: Digitalized Setting for Time-Delayed Quantitative Control Actuator: With foam alarm function |
9. | Supplement materials | Control Method: Manual/Automatic Control Content Control: Digitally Set Segmented Quantitative Control Actuator: Peristaltic pump control or solenoid valve control |