Product Introduction
This product is a laboratory-standard fully automatic fermentation tank, capable of meeting general testing requirements. It is widely used and features a comprehensive range of functions.
1.1 System Composition
Serial Number | System | Dimensions (L) | Tank Material | Saddle Jacket / Pipe Bundle Material | Quantity | Note |
1. | Fermenter | 20 | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Functions
Serial Number | System/Equipment | Can Lid Lifting and lowering | Control Function | |||||||
Temperature | Ph | DO | RPM | Traffic | Pressure | Sterilization | Supplementary Material | |||
1. | Fermenter | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Self-contained | Self-contained | Automatic | Solenoid Pumps |
1.3. Control Parameters
Serial Number | Control Parameters | Technical Specifications |
1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode accessories Measurement Accuracy: 0.3 grade (0.3% FS ± 0.1℃), Measurement Range 0-150℃ Control Method: Digital settings, manual/automatic control Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps Content Control: Bi-directional Control of Heating and Cooling |
2. | PH | Swiss Company PH Glass Gel Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Stainless Steel PH Electrode Sheath with Matching Components Measurement Range: 0-14PH, Accuracy: ±0.01PH Four-digit display transmitter, powered by AC 220V, 4-20mA Original Swiss-imported electrode shielding wire equipped |
3. | DO | Swiss Company DO Membrane Electrode Electrodes can be sterilized in situ at high temperatures, sterilization temperature: 130°C. Matching stainless steel DO electrode sheath. Measurement range 0-100.% with ±1% accuracy. Supplied with a four-digit display transmitter, powered by AC 220V, 4-20mA |
4. | RPM | Variable Frequency Continuous Speed Control System, power matched with motor Digitized setting of revolution value Actuator: Variable Frequency Drive |
5. | Sterilization | Digitize the setting of sterilization temperature and time inside the tank, with automatic control of the sterilization and cooling process. Sterilization temperature range: 80℃ - 125℃. Sterilization Process: Steam is passed through the jacket to heat the drum internally; once the preheating set temperature is reached, you can choose to set the steam to enter the drum directly. Achieve sterilization temperature inside the can; start timing and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. |
6. | Pressure | Digitally set the tank pressure; linearly adjust tank pressure with a 4-20mA signal. |
7. | Traffic (optional) | Companion quality flow meters control the inflow of air into the tank |
8. | Defoamer | Control Methods: Manual/Automatic Control Content Control: Digital Setting for Time-Delayed Quantitative Control Actuator: Equipped with foam alarm function |
9. | Supplement Materials | Control Method: Manual/Automatic Content Control: Digitized Stepwise Quantitative Control Actuator: Peristaltic pump control or solenoid valve control |