Product Introduction
This product is a standard laboratory-type fully automatic fermenter, capable of meeting general testing requirements. It is widely used and features a comprehensive range of functions.
1.1 System Composition
Serial Number | System | Specs (L) | Vessel Material | Spiral Jacket/Multi-tube Material | Quantity | Note |
1. | Fermenter | 10 | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Functions
Serial Number | System/Equipment | Can lidLifting and lowering | Control Function | |||||||
Temperature | Ph | DO | RPM | Traffic | Pressure | Sterilization | Supplementary Material | |||
1. | Fermenter | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Self-controlled | Self-controlled | Automatic | Solenoid Pump |
1.3. Control Parameters
Serial Number | Control Parameters | Technical Specifications |
1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode included Measurement Accuracy: 0.3 grade (0.3% FS ± 0.1℃), Measurement Range 0-150℃ Control Method: Digital setting, manual/automatic control Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps Content Control: Bi-directional Control of Heating and Cooling |
2. | PH | Swiss Company PH Glass Gel Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Stainless Steel pH Electrode Sheath; Measurement Range: 0-14PH, Accuracy: ±0.01PH Four-digit display transmitter, power supply AC 220V, 4-20mA Original Swiss进口 electrode shielding wire equipped |
3. | DO | Swiss Company DO Membrane Electrode Electrodes can be sterilized in situ at high temperatures, sterilization temperature: 130°C. Stainless steel DO electrode sheath with accessories. Measurement Range: 0-100.%; Accuracy: ±1% Supplied with a four-digit display transmitter, powered by AC 220V, 4-20mA. |
4. | RPM | Variable frequency stepless speed control system, power matched with the motor Digitalized setting for rotation speed values Actuator: Variable Frequency Drive |
5. | Sterilization | Digitally set the sterilization temperature and time inside the tank, with automatic control over the sterilization temperature and cooling process. Sterilization temperature range: 80℃-125℃. Sterilization Process: Steam is heated through a jacket to preheat the tank; once the preheat set temperature is reached, you can opt to set steam to directly enter the tank. Achieve sterilization temperature inside the can; start timing and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. |
6. | Pressure | Digitally set the tank pressure, linearly adjust the tank pressure with a 4-20mA signal. |
7. | Traffic (optional) | Flowmeter-assisted control of air flow into the tank |
8. | Defoamer | Control Method: Manual/Automatic Control Control Content: Digitalized Setting for Time-Delayed Quantitative Control Actuator: Foam Alarm Function |
9. | Supplement Materials | Control Method: Manual/Automatic Control Content Control: Digitized Stepwise Quantitative Control Actuator: Peristaltic pump control or self-regulating valve control |