Product Introduction
This product is a standard laboratory fully automatic fermentation tank, capable of meeting general testing requirements. It is widely used and feature-rich.
1.1 System Composition
Serial Number | System | Specification (L) | Tank material | SHELL/PIPE TUBE MATERIAL | Quantity | Note |
1. | Fermenter | 30 | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Function
Serial Number | System/Equipment | Can lid Lifting and lowering | Control Function | |||||||
Temperature | Ph | DO | RPM | Traffic | Pressure | Sterilization | Supplementary Materials | |||
1. | Fermentation Vessel | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Self-contained | Self-regulating | Automatic | Solenoid Pumps |
1.3. Control Parameters
Serial Number | Control Parameters | Technical Specifications |
1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode compatible Measurement Accuracy: 0.3 grade (0.3%FS ± 0.1℃), Measurement Range 0-150℃ Control Method: Digital setting, manual/automatic control Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps Content Control: Bidirectional Heating and Cooling Control |
2. | PH | Swiss Company PH Glass Gel Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Stainless Steel pH Electrode Sheath with Matching Accessories Measurement Range: 0-14PH, Accuracy: ±0.01PH Four-digit display transmitter, power supply AC 220V, 4-20mA Original Swiss进口 electrode shielding wire equipped |
3. | DO | Swiss Company DO Membrane Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Matching stainless steel DO electrode sheath. Measurement range: 0-100.%; Accuracy: ±1%; Supplied with a four-digit display transmitter, powered by AC 220V, 4-20mA. |
4. | RPM | Variable Frequency Stepless Speed Control System, matched with motor power Digitalized setting for speed values Actuator: Variable Frequency Drive |
5. | Sterilization | Digitally set the sterilization temperature and time inside the tank, with automatic control over the sterilization temperature and cooling process. Sterilization temperature range: 80℃-125℃. Sterilization Process: Steam is passed through the jacket to heat the tank internally; once the preheating set temperature is reached, you can choose to set the steam to enter the tank directly. Achieve sterilization temperature inside the can; start timing and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. |
6. | Pressure | Digitize the setting of the pressure inside the tank.4-20mA signal linearly adjusts the tank's internal pressure. |
7. | Traffic (Optional) | Matching quality flowmeter to control the air flow into the tank |
8. | De-foaming | Control Methods: Manual/Automatic Control Control Content: Digitalized Time Delay Gravimetric Control Actuator: Equipped with foam alarm function |
9. | Supplementary Material | Control Method: Manual/Automatic Control Content Control: Digitized Segmented Quantitative Control Actuating Mechanism: Peristaltic Pump Control or Self-Control Valve Control |