Product Introduction
This product is a standard laboratory-type fully automatic fermenter, capable of meeting general testing requirements. It is widely used and features comprehensive functionality.
1.1 System Composition
Serial Number | System | Dimensions (L) | Tank material | Saddle/Coiled Tube Material | Quantity | Note |
1. | Fermenter | 100 | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Functions
Serial Number | System/Equipment | Can lidLifting & Lowering | Control Function | |||||||
Temperature | Ph | DO | RPM | Traffic | Pressure | Sterilization | Supplement materials | |||
1. | Fermentation Vessel | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Manual | Inspection | Automatic | Solenoid Pumps |
1.3. Control Parameters
Serial Number | Control Parameters | Technical Specifications |
1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode accessory Measurement Accuracy: 0.3 grade (0.3%FS ± 0.1℃), Measurement Range: 0-150℃ Control Method: Digital Setting, Manual/Automatic Control Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps Content Control: Bidirectional Control of Heating and Cooling |
2. | PH | Swiss Company PH Glass Gel Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Stainless Steel pH Electrode Sheath for Matching Measurement Range: 0-14PH, Accuracy: ±0.01PH Four-digit display transmitter, powered by AC 220V, 4-20mA Original Swiss进口 electrode shielding wire equipped |
3. | DO | Swiss company DO membrane electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Matching stainless steel DO electrode sheath Measurement range: 0-100%, accuracy ±1% Supplied with a four-digit display transmitter, powered by AC 220V, 4-20mA. |
4. | RPM | Variable frequency stepless speed control system, power matched with motor Digitalized setting of speed values Actuator: Variable Frequency Drive |
5. | Sterilization | Digitize the setting of sterilization temperature and time inside the tank, and automatically control the sterilization and cooling process. Sterilization temperature range: 80℃-125℃. Sterilization Process: Steam is passed through the jacket to heat the tank internally; once the preheating set temperature is reached, you can opt to set the steam to enter the tank directly. Achieve sterilization temperature inside the can; start timing and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. |
6. | Pressure | Manual control |
7. | Traffic | Glass Rotameter |
8. | Defoaming | Control Method: Manual/Automatic Control Content Control: Digital Set Time Quantity Control Actuator: With foam alarm function |
9. | Supplement Materials | Control Method: Manual/Automatic Control Content Control: Digitized Segmental Quantitative Control Actuating Mechanism: Peristaltic Pump Control or Auto-Valve Control |