Product Introduction
This product is a standard laboratory全自动fermenter, capable of meeting general testing requirements. It boasts wide applicability and comprehensive functions.
1.1 System Composition
Serial Number | System | Dimensions (L) | Vessel Material | Sleeve/Tube Bundle Material | Quantity | Note |
1. | Fermenter | 50 | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Function
Serial Number | System/Equipment | Can lidLifting and lowering | Control Function | |||||||
Temperature | Ph | DO | RPM | Traffic | Pressure | Sterilization | Supplementary materials | |||
1. | Fermenter | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Self-controlled | Self-controlled | Automatic | Solenoid Pump |
1.3. Control Parameters
Serial Number | Control Parameters | Technical Specifications |
1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode included Measurement Accuracy: 0.3 grade (0.3% FS ± 0.1℃), Measurement Range 0-150℃ Control Method: Digital settings, manual/automatic control Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps Content Control: Bidirectional Control of Heating and Cooling |
2. | PH | Swiss Company PH Glass Gel Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Stainless Steel pH Electrode Sheath for Matching Measurement Range: 0-14PH, Accuracy: ±0.01PH Four-digit display transmitter, powered by AC 220V, 4-20mA Original Swiss进口 electrode shielding wire equipped |
3. | DO | Swiss Company DO Membrane Electrode Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. Matching stainless steel DO electrode sheath Measurement range: 0-100%, accuracy ±1% Supplied with a four-digit display transmitter, power supply AC 220V, 4-20mA |
4. | RPM | Variable-frequency stepless speed regulation system, power matched with motor Digitalized setting of rotation speed values Actuator: Variable Frequency Drive |
5. | Sterilization | Digitize the setting of the pot's sterilization temperature and time, with automatic control over the sterilization temperature and cooling process. Sterilization temperature range: 80℃-125℃. Sterilization Process: Steam is heated through the jacket to preheat the tank; once the preheat set temperature is reached, you can choose to set the steam to directly enter the tank. Achieve sterilization temperature inside the can; start timing and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. |
6. | Pressure | Digitize the setting of the tank pressure.The 4-20mA signal linearly adjusts the tank's internal pressure. |
7. | Traffic (optional) | Kit quality flowmeter to control air flow into the tank |
8. | De-foaming | Control Method: Manual/Automatic Control Content Control: Digital Setting for Time-Delayed Gravimetric Control Actuator: With foam alarm function |
9. | Supplementary materials | Control Method: Manual/Automatic Control Content Control: Digitized Segmental Quantitative Control Actuator: Peristaltic pump control or self-regulating valve control |