Product Description
This product is a standard laboratory full-automatic fermentation tank, capable of meeting general testing requirements. It is widely used and boasts comprehensive functions.
1.1 System Composition
| Serial Number | System | Dimensions (L) | Drum Material | Suction Sleeve/Metallic Tube Material | Quantity | Note |
| 1. | Fermenter | 30 | SUS316L | SUS304 | 1 | Operates independently |
1.2 Control Function
| Serial Number | System/Equipment | Can lid | Control Function | |||||||
| Lifting & Lowering | Temperature | Ph | DO | RPM | Traffic | Pressure | Sterilization | Supplementary Materials | ||
| 1. | Fermenter | Automatic | Measurement and Control | Measurement and Control | Measurement and Control | Variable Frequency | Manual | Inspection | Automatic | Solenoid Pumps |
1.3 Control Parameters
| Serial Number | Control Parameters | Technical Specifications |
| 1. | Temperature | Four-digit temperature transmitter, with PT100 temperature electrode kit |
| Measurement Accuracy: 0.3 grade (0.3% FS ± 0.1℃), Measurement Range: 0-150℃ | ||
| Control Method: Digital setting, manual/automatic control | ||
| Actuating Mechanism: Pneumatic Valves/Hot Water Circulation Pumps | ||
| Content Control: Bidirectional Control of Heating and Cooling | ||
| 2. | PH | Swiss Company PH Gel Electrode for Glass |
| Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. | ||
| Stainless steel pH electrode sheath with accessories. | ||
| Measurement Range: 0-14PH, Accuracy: ±0.01PH | ||
| Four-digit display transmitter, power supply AC 220V, 4-20mA | ||
| Original Swiss进口 electrode shielding wire equipped | ||
| 3. | DO | Swiss company DO membrane electrode |
| Electrodes can be sterilized in place at high temperatures, sterilization temperature: 130°C. | ||
| Matching stainless steel DO electrode sheath. | ||
| Measurement range: 0-100%, accuracy: ±1% | ||
| Supplied with a four-digit display transmitter, powered by AC 220V, 4-20mA. | ||
| 4. | RPM | Variable frequency stepless speed control system, power matched with the motor |
| Digitalized setting of speed values | ||
| Actuator: Variable Frequency Drive | ||
| 5. | Sterilization | Digitize the setting of the sterilization temperature and time inside the tank, with automatic control over the sterilization and cooling processes. |
| Sterilization temperature range: 80℃-125℃. | ||
| Sterilization Process: Steam is heated through the jacket to preheat the tank; once the preheat set temperature is reached, you can opt to set the steam to enter the tank directly. | ||
| Achieve sterilization temperature inside the can; start timing and automatically control the sterilization temperature; after reaching the sterilization time, automatically cool down to the set fermentation temperature. | ||
| 6. | Pressure | Manual control |
| 7. | Traffic | Glass Rotameter |
| 8. | Defoaming | Control Method: Manual/Automatic Control |
| Content Control: Digital Setting for Time-Delayed Quantitative Control | ||
| Actuator: with foam alarm function | ||
| 9. | Supplementary Material | Control Method: Manual/Automatic Control |
| Content Control: Digitized Segmental Quantitative Control | ||
| Actuator: Peristaltic pump control or self-regulating valve control |