Overview
Built-in with Windows 10 system, this 12-inch industrial tablet PC can directly perform differential scanning calorimetry (DSC) operations without the need for additional computer connection, generate test reports, and connect to a printer via USB. As a classic thermal analysis method for studying the thermal effects at controlled temperatures, DSC has been widely applied in various fields such as research and development, process optimization, quality inspection and control, and failure analysis of materials and chemicals today. Using the DSC method, we can study phase transitions of inorganic materials, melting and crystallization processes of polymer materials, polymorphism of drugs, solid/liquid phase ratios of food oils, and more. Compliant with national standards GB/T2951.42-2008, GB/T15065-2009, GB/T17391-1998, and GB/T19466.6-2009.
Technical Features
1. Built-in with Windows 10 system, this 12-inch industrial tablet PC allows direct completion of differential thermal testing without the need to connect to a computer.
2. Connect via RS23 or USB cable to the computer.
3. The furnace body features a compact structure with adjustable heating and cooling rates.
4. Enhanced installation techniques, utilizing full mechanical fixation, to completely avoid contamination of differential thermal signals by the internal gel of the furnace.
5. Dual-temperature probe ensures high repeatability in sample temperature measurement.
6. The digital gas mass flow meter automatically switches between two gas flow paths, with fast switching speed and short stabilization time.
7. Standard sample included for customer convenience in adjusting temperature coefficients.
8. Generate test reports directly and connect to a printer to print them.
9. Supports user-programmable routines for full automation of measurement steps. The software offers dozens of commands, allowing users to flexibly combine various instructions based on their measurement procedures and save them. Complex operations are simplified into one-click operations.
10. Optional refrigeration module (suitable for materials with low glass transition or melting points, or for rapid cooling requirements).
Technical Specifications
1. DSC Range: 0 ~ ±250mW
2. Temperature Range: Room temperature to 500°C
3. Heating Rate: 0.1~80℃/min
4. Temperature Resolution: 0.01℃
5. Temperature Accuracy: ±0.1℃
6. Temperature Repeatability: ±0.1℃
7. DSC Accuracy: ±2%
8. DSC Resolution: 0.001mW
9. DSC Resolution: 0.001mW
10. Temperature Control Methods: Heating, Constant Temperature, Cooling, and Circulating Temperature Control (fully automatic program control)
11. Curve Scanning: Heating Scan
12. Atmosphere Control: The gas mass flow meter automatically switches between two gas lines
13. Display: 24-bit color, 12-inch touchscreen display
14. Data Interface: USB standard interface, with corresponding operation software
15. Parameter Standards: Equipped with standard calibration materials and one-touch calibration function, allowing users to calibrate temperature independently.
16. Operating Power: AC 220V, 50/60Hz
Application Examples
Measurements of physical and chemical changes related to heat, such as glass transition temperature, melting point, fusion temperature, crystallization and crystallization heat, phase transition reaction heat, product thermal stability, curing/cross-linking, oxidation induction time, reaction kinetics, specific heat, etc.