Oil Particle Counter, based on standards such as GB/T 18854-2002 (ISO 11171-1999). Utilizes the light-blocking counting principle, fully compliant with relevant military, national, and international standards. Offers fast, accurate, reliable, and reproducible test results along with comprehensive contamination monitoring and analysis reports. Suitable for particle contamination detection in hydraulic oils, lubricating oils, anti-knock fuels, insulating oils, and turbine oils. Widely applicable in industries such as aviation, aerospace, power, petrochemicals, transportation, ports, metallurgy, machinery, and automotive manufacturing.
Key Features
Utilizing the light-blocking (dark-field) counting principle established by the International Committee for Weights and Measures.
High-precision laser sensors with wide testing range, stable performance, low noise, and high resolution.
High-pressure syringe pump sampling method, allows for self-setting of sample volume, stable feeding speed, and high sampling accuracy.
A combined positive-negative pressure sampling system that achieves degassing of samples, suitable for testing products of varying viscosities.
Internally equipped with a pressure sensor, capable of setting pressure values and automatically determining the cabin's air pressure to ensure safety.
Built-in air purification system ensures testing is not contaminated
Internally equipped with multiple calibration curves, compatible with all commonly used domestic and international standards for calibration.
Incorporates various commonly used standards including GJB-420A, GJB-420B, NAS1638, GB/T14039, ISO4406, SAE4059cpc, SAE4059F, SAE749D, ГOCT17216, QC/T29104, JB/T9737, DLT432, HH005-2018, supports customized standard testing, and can be configured to meet specific customer standards as required.
Built-in data analysis system, one test provides all standard test data and contamination levels
Integrated viscosity, moisture, and temperature sensor modules, providing precise particle distribution testing along with viscosity, moisture content saturation, ppm values, and temperature reference (optional).
Size particles to your preference, with an in-built selection of nearly ten thousand particle sizes for easy particle size analysis.
Multiple sampling containers, such as standard sampling bottles or cups, are available to meet the testing requirements of various industries.
Full-function automatic color touch screen operation, supporting dual Chinese and English interfaces, easy and convenient to operate
Single-channel and multi-channel calibration capabilities for automatic calibration function
Equipped with an RS232 interface for connection to a computer or laboratory platform for data processing
With extensive data storage and printing capabilities, it can store up to 1000 sets of data and supports USB flash drive data storage.
Technical Specifications
Light Source: Semiconductor Laser
Particle Size Range: 0.8μm ~ 600μm (calibrated in 1-100μm or 4-70μm (c))
Detection Channels: 8-64 channels available, particle size can be set arbitrarily
Sensitivity: 0.8 μm (ISO 4402) or 3 μm (c) (GB/T 18854, ISO 11171)
Resolution: <10% (GB/T18854, ISO 11171)
Repeatability: RSD < 2%
Sample viscosity test: ≤650cSt (viscosity too high? Use heating or dilution methods for testing)
Sample Volume: 0.2~1000mL, in 0.1mL increments
Sampling Accuracy: Better than ±0.5%
Sampling Rate: 5 ~ 80 mL/min
Lug Pressure Vessel: Positive and negative pressure vessel setup for degassing of samples and testing of high viscosity samples
Maximum Vacuum of Pressure Compartment: -0.08 Mpa
Maximum positive pressure in pressure cabin: 0.8 Mpa
Maximum particle concentration: 12,000~40,000 particles/mL
Temperature (Optional): Range: 1~100℃; Accuracy: 1℃
Water Activity (optional): Collection Range: 1~100%RH; Collection Accuracy: 1%RH
Moisture Content (optional): Collection Range: 1~300ppm; Collection Accuracy: 1ppm
Power: 110~245V AC, 50Hz, 70W
Dimensions: 340mm x 410mm x 650mm
Instrument Net Weight: 23 kg