Fully Automatic Glove Box ZKYQX-I Gas Purification and Circulation System
One. Glove Box Function
Sealed Recirculation
The inert gas within the glove box is circulated and purified through a closed loop by a recirculating fan and purifier, continuously removing moisture and oxygen.
Automatic Control Regeneration
Deoxygenation materials can be regenerated, with the regeneration process controlled by a program.
Cabinet Pressure Control
The glove box's internal pressure is automatically controlled by a PLC. The working pressure can be freely set within +/- 1500Pa, and the system automatically protects against deviations beyond +/- 1500Pa.
Vacuum pump automatic control
Vacuum pumps are required to automatically activate when needed in the system.
II. Main Components and Configurations
Integrated with a sealed housing (including a sloped operating surface safety glass one-way vision window), transition compartment, recirculation purification unit, PLC control and touch screen operating system, vacuum system, display system, bracket (integral) etc.
Basic Configuration
1 stainless steel 304 cabinet
One large transition compartment made of stainless steel
1 mini transition cabin
1 pair of butyl rubber gloves
1 lighting system
1 piece per box power interface
4 KF40 spare interfaces
1 high-performance recirculating fan, 90 m³/h capacity
1 vacuum pump
1 set of water probe
1 Oxygen Probe Set
Stool casters
Foot switch
III. Technical Specifications:
Water Oxygen Index: Less than 0.1 ppm.
2. Leak Rate: <0.001 vol%/h
3. Technical Specifications
1). Glove box housing
(1) Case: 304 stainless steel, thickness 3mm
(2) Inner surface: Stainless steel; Outer surface: Powder coated, white
(3) Studio Size: 1220mm x 760mm x 900mm (L x W x H)
(4) Dimensions: 1935mm x 790mm x 1800mm (Length x Width x Height)
(5) Front Window: Inclined view window, 8mm thick transparent tempered safety glass, coated with anti-corrosion film, wear-resistant, corrosion-resistant, and good light transmission.
(6) Glove Mouth: Made of anodized aluminum, sealed with an O-ring, diameter 220mm
(7) Gloves: Butyl rubber, thickness 0.4mm, diameter 8", length 32"
(8) Filter: 0.3 micron specification, 1 gas inlet and 1 gas outlet
(9) Shelves: Made of stainless steel, adjustable height.
(10) Cabinet Lighting: LED lights, installed above each glass window in front
(11) Interface: 4 spare interfaces, DN 40 KF material: SUS304 stainless steel; chuck-type quick assembly interface, allowing gas or liquid to enter the box safely and conveniently.
(12) Power Interface: 1 (220V) located inside the glove box work compartment, providing power to electrical equipment. A power strip is installed inside the glove box, featuring three sets of outlets with both two-prong and three-prong options.
(13) Foot switch: Controls the gas replenishment and evacuation of the housing, located below the housing, for fine-tuning the pressure inside the work chamber.
(14) Glove Box Frame: Carbon steel with electrostatic powder coating, one-piece design; equipped with swivel casters and adjustable support legs; portable glove box.
(15) Vacuum Gauge: Mechanical dial display, measures vacuum levels in overpressure and tool overpressure chambers, WIKA Germany.
4. Large Transition Hatch
On the right side of the housing; diameter 385mm x 600mm, material: SUS304 stainless steel;
Airtightness: Leakage rate ≤ 0.001 Vol%/h.
(1) Attachment
Slide Trays: 304 Stainless Steel
Hatch: Double door, anodized aluminum material, thickness 10mm, vertical operation, equipped with a pressure gauge with analog display
(2) Control
Electromagnetic Valve Touchscreen Automatic Operation
5. Small transition compartment
Size: Diameter 150mm, Length 300mm, Length into the housing 100mm
Material: 304 Stainless Steel
Surface: Inner surface with brushed finish, outer surface painted (white)
(1) Attachment
Hatch: Double doors, pressure-latch opening method Pressure gauge: Analog display
(2) Control
Manual Valve - Manual Operation
6. Gas Purification and Recirculation System
Purification Column
Function: Gas-tight, moisture and oxygen removal (recirculation capacity 90 m³/h)
Container Material: 304 Stainless Steel
Purification Materials: Copper Catalyst: 4kg; Molecular Sieve: 6kg
Purification Capacity: Oxygen Removal: 60L; Water Removal: 2Kg
Water Oxygen Index: Less than 0.1ppm (empty box)
(2) Circulatory System
Working Gases: Nitrogen, Argon
Recirculation Capacity: Integrated Fan Flow Rate 90 m³/h
(3) Recycled
Operation: PLC automated control of regeneration process
Argon Gas: working gas mixed with H2 (5-10% H2)
(4) Vacuum Pump
Spec: 12m3/h, scroll pump, with oil mist filter, air vibration control
Valve
Main Valve: DN 40 KF, Electromagnetic Angle Valve
Control Valve: Electromagnetic Integrated Valve
7. Control System
Function: Includes self-diagnosis, power-off self-start feature, with pressure control and adaptive functions; automatic control, cycle control, password protection; unit control with Siemens PLC touch screen.
Pressure Control: The pressure in the housing and transition compartment can be freely set within +/- 1500 Pa working pressure, with automatic protection activated beyond +/- 1500 Pa.
Foot pedal: Controls the pressure inside the cabinet, facilitating the operation of increasing and decreasing pressure.
Display System: Utilizes Siemens PLC touch screen PT177 to show operational status, cabinet pressure, system records, etc.
8. Vacuum System Control Status
Vacuum Pump - Flow rate: 8 m³/h, Ultimate Vacuum: ≤2x10^-1 Pa. Can be started manually or via PLC, suitable for vacuuming transition chambers and maintaining cabinet pressure balance. Ultimate Vacuum: ≤2x10^-1 Pa.
9. Oxygen probe
Imported Brand AII from the USA, featuring ±0.1ppm high precision, electrochemical sensor, 0-1000PPM display values, and low drift.
10. Water Probe
Imported Vaisala (Finland), measurement range: 0-1000PPM with clear readings, low drift, wide application, especially for lithium battery manufacturing and metal-organic users. Can be restored to initial state through cleaning and regeneration process, avoiding the issue of one-time contamination leading to disposal.
Water and Oxygen Indices: Measurement data from the water and oxygen analyzers must be displayed, and an alarm must sound if values exceed the set limits.
11. Acceptance Criteria
Under standard conditions, at a standard atmospheric pressure, using a 99.999% pure inert gas source, and with the container running empty, the water oxygen content is less than 0.1 ppm, and the container leakage rate is ≤0.001 Vol%/h.