Compass 4294 Plus
Portable oil sulfur-chlorine analysis spectrometer
Simply the Best
Ø Simultaneous non-destructive testing for sulfur and chlorine elements in oil products
Ø Portable integrated design meets stringent on-site testing requirements
Ø Vacuum testing environment, effectively enhances sensitivity
Ø Ultra-low detection limit
Ø One-touch start and one-touch test
Ø Easily meet the following standards for oil quality control, in compliance with the following specifications:
National Standard
GB/T 17040-2008 Sulfur Content Determination in Chinese Petroleum and Petroleum Products
GB/T 380-1977 Sulfur Content Measurement of Chinese Petroleum Products
GB/T 17411-2015 Marine Fuel Oil
SH/T 0253-1992 Total sulfur content determination in light petroleum products
ASTMAndISOInternational Standards
ASTM D4294, ISO 8754, ISO 20847, IP 336, ASTM D6445,IP496, ASTM7220
ASTM D4929
Background
Sulfur is a harmful substance in oil.,Sulfide of iron, ferric sulfate, sulfite, or sulfuric acid may easily form.,Severe corrosion of equipment. The combustion of gasoline also leads to significant environmental pollution, hence it is crucial to continuously reduce the sulfur concentration in fuel.
Chlorides are always present in oil, with concentrations varying greatly. Organic chlorines decompose during processing, leading to severe corrosion incidents throughout the refinery.EDXRFDetermining sulfur and chlorine levels in oil or fuel is one of the standard industrial analytical methods in modern times.
In China,2018.11.30The Ministry of Transport issued the "Implementation Plan for the Control Area of Ship Air Pollutant Emission," effective as of2022Year1Month1From the date on.,Vessels entering the coastal control area of Hainan waters should use fuel with sulfur content not exceeding0.1%m/mMarine Fuel 。
The new national V standard for automotive gasoline and diesel, as well as the VI standard, requires the allowable sulfur content to be less than10 ppm。

Introduction
Compass 4294 plusEnergy DispersiveXRay FluorescenceEDXRFWe offer a novel method for analyzing sulfur and chlorine content in petroleum products on-site, with a concentration range of2 ppmTo10Percentages vary.
To enhance detection sensitivity,4294 plusUtilized withPeltierHigh-resolution, large-area, rapid silicon drift detector with advanced cooling functionFSDDDetector. FSDDSpectral resolution of the detectorFWHM) Less than<130eV(MnKα), Input counting rate can reach up to500,000 cps。
Features and Advantages
•Portable, sturdy, and compact design for effortless on-site high-precision sample non-destructive testing
•Vacuum detection chamber, enhancing sulfur and chlorine detection sensitivity3Multiplied by several times
•Ultra-high resolution detector
•Special optical path design, enhancing detection lower limit
•User-friendly software interface and one-click test launch
•Standard analysis training takes only a few minutes.
•Low-cost routine analysis, no gas consumption required
•Large8Intuitive interface displayed on an inch touchscreen
•Optional for high-capacity battery, enabling long working time without external power supply
Application Fields
The petrochemical field encompasses sulfur content analysis of gasoline, kerosene, and other fuels.
Sulfur Content Analysis of Marine Vessel Fuels
Marine fuelCl,MgAndKAnalysis
Other fields requiring on-site element analysis
Technical Specifications
Detector | High resolutionFSDDDetector |
Exciting Source | 4Washable mini integrated light tube 50kv Max,200uA Max |
Operating Temperature | -20 to 50 °C |
Vacuum state | Vacuum(4294 PlusBuilt-in vacuum pump) |
Measurement Time | 130-300Seconds (user-defined test duration) |
Sample Type | Liquids, solids, and powders |
Instrument Calibration | Auto-calibrate upon startup |
Calibration Curve | Low-sulfur: 0ppm- 100ppm High-sulfur: 0.01%-5% |
Instrument dimensions | 270mm*320mm*230mm(Long*Wide*High) |
Sample cavity size | 170mm*110mm*17mm(Long*Wide*High) |
Instrument weight | 10.5kg |
Display Screen(Resolution1280*800) | 8inches Win10Operating System |
Data Transmission | USB Port, Bluetooth, Wi-Fi, GPS |
Battery (Optional) | Work for eight consecutive hours |
Safety Protection | Radiation Indicator Light Built-in radiation protection device(Automatic Cut Off) Software Alert |
Report Format | Excel, PDF |
Accessories and consumables | Test Film Oil Test Cup Sample Tube Vacuum Test Window Film |
Achieve high-precision results without daily calibration. The instrument comes with a factory-calibrated curve, allowing customers to directly measure samples.
Continuous Test Report for Sulfur and Chlorine in Oil Standard Sample
Measurement Frequency. | Calibration Curve | SulfurS-50ppm | SulfurS-10ppm | SulfurS-100ppm | SulfurCl-100ppm |
|
1 | Fuel Oil | 50.64 | 11.25 | 104.57 | 96.5 | |
2 | Fuel Oil | 50.26 | 11.84 | 104.55 | 97.1 | |
3 | Fuel Oil | 49.96 | 10.34 | 106.27 | 98 | |
4 | Fuel Oil | 52.75 | 11.9 | 102.31 | 95.7 | |
5 | Fuel Oil | 50.28 | 9.73 | 107.04 | 99 | |
6 | Fuel Oil | 48.20 | 10.64 | 102.94 | 102.1 | |
7 | Fuel Oil | 49.90 | 10.49 | 96.35 | 99.9 | |
8 | Fuel Oil | 51.18 | 10.49 | 101.84 | 96.1 | |
9 | Fuel Oil | 52.99 | 11.03 | 101.19 | 101 | |
10 | Fuel Oil | 52.61 | 8.81 | 101.18 | 102.6 |
|
Average measurementAverage (ppm) | 50.88 | 10.65 | 102.82 | 98.80 | Element | |
Standard DeviationSD(%) | 1.443 | 0.887 | 2.902 | 2.390 | ||
3Multiple Standard Deviation3s | 4.329 | 2.660 | 8.706 | 7.171 | SulfurSulfur- S ChlorineChlorine- Cl | |
Relative Standard DeviationRSD | 2.836% | 8.324% | 2.822% | 2.419% | ||
Actual sulfur content of the sample Certified Value(ppm) | 50.00 | 10.00 | 100.00 | 100.00 | ||
Sulfur sensitivity enhanced under vacuum conditions3Multiple times
>>The purple spectrum was obtained under air measurement conditions, and the red spectrum was acquired through4294 plusObtained in a vacuum measurement environment