Shenzhen Zhongnuo Testing Technology Co., Ltd. specializes in the calibration, verification, and metrological services for various gas alarms, flammable gas detectors, and toxic gas monitors, catering to all types of locations. We hold the necessary qualifications to serve nationwide, addressing clients' electrical safety and explosion-proof fire prevention issues, and contributing to safe production.
Gas Alarm Meters Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
2. Relative Humidity: ≤85%
3. The working environment should not interfere with the normal operation of the instrument by electromagnetic fields or interfering gases.
Calibration Tools and Instruments
1) Gas Standard Material
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Device: Dilution error of 1%.
3) Stopwatch: Graduation not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that matches with the calibrated cylinder.
7) Gas Pipeline: Tubing made of materials like polytetrafluoroethylene or stainless steel that do not affect the concentration of the tested gas.
Measurement Equipment Control
Instrument control includes calibration, follow-up calibration, and in-service inspection.
Which gas detectors require regular calibration and metrological verification?
Fixed-install, handheld, and portable gas detectors, gas alarm instruments, toxic and harmful gas detectors/alarm systems, flammable gas detectors/alarm systems, gas analyzers, etc.
How often should various gas detectors be periodically calibrated and verified?
The recalibration period for instruments is typically one year, which is determined by the sending unit based on actual usage. During the re-calibration process, if there are any doubts about the instrument's test data, or if the main components of the instrument are replaced, as well as after repairs, a recalibration is required.
Shenzhen ZhongNuo Inspection Technology Co., Ltd. is a third-party institution that offers regular certification, measurement, and calibration services for electrical safety, electrical explosion-proofing, gas alarms, and toxic/hazardous gas detection instruments, with qualifications to address fire and explosion prevention issues for clients.