Product Application:
Equipment for environmental protection, hygiene and epidemic prevention, pharmaceutical inspection, agriculture and sideline products, and aquatic products research, schools, and production departments. It is used for water analysis, BOD measurement, various constant-temperature tests, environmental tests, cultivation, preservation of bacteria, molds, microorganisms, and plant cultivation and breeding experiments.
Product Features:
1. Equipped with a compressor set protected by high and low pressure, intelligent defrost-free refrigeration technology, self-diagnostic compressor on/off control program, low noise, long lifespan, and minimal temperature fluctuations.
2. High-performance CPU processing chip and highly sensitive, liquid crystal display. The high-precision platinum resistance sensor-based temperature control system ensures more precise temperature control and easier operation.
3. Equipped with sensor failure alarm, over-temperature alarm, self-diagnostic dynamic control, temperature display calibration, parameter memory, and a timing function up to 9999 minutes.
4. Independent temperature control and alarm system; automatically shuts off operation upon exceeding set temperature; double safety assurance.
Operation Procedure:
Normal operating conditions
Relative Humidity: 80% RH
Ambient Temperature: 5-32℃
Pressure: 86-106 Kpa
The incubator is surrounded by no intense vibrations or corrosive gases.
The incubator should be shielded from direct sunlight or any other heat/cold sources.
Usage Instructions:
Connect the power supply to the instrument as specified and ensure the grounding terminal of the power supply socket is reliably grounded.
2 Turn on the power switch, the power indicator light illuminates, and the temperature controller begins to display the indoor temperature.
Set temperature according to the needs of the culture, and after achieving the set temperature, it will automatically maintain it.
Important Notes and Routine Maintenance:
Firstly, ensure proper grounding of the equipment, using a wire twice as thick as the power cable for grounding.
2. Samples should not be placed at the bottom of the studio, and the samples on the shelves should not be too crowded to avoid affecting the air circulation inside the box.
3. When resuming use after long-term storage, the equipment should be started for 6-8 hours without load to clear any potential issues that may have occurred during transportation or storage, before normal operation to avoid unnecessary losses.
4. The biohazard box must be placed in a location that is shaded, well-ventilated, and dry.
5. Do not use acids, alkalis, or other corrosive substances to clean the inner and outer surfaces. Use a neutral detergent for regular cleaning and wipe dry with a soft cloth.
6. When replacing a fuse, replace it with a fuse of the same current rating and size as indicated by the fuse markings.
7. When the biohazard box is not in use, the power should be disconnected and the box should be kept clean and dry.
Product Specifications: