Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and metrological services for gas alarms, combustible gas detectors, and toxic/hazardous gas detectors for various locations. We are fully certified to serve nationwide, providing solutions for all types of electrical safety and explosion-proof fire prevention issues, contributing to safe production.
Power-on inspection
The instrument should operate normally after being powered on, with clear and complete display sections. Check the alarm function and the alarm action values.
Alarm function and alarm action value verification
The instrument's sound and light alarm should be functioning normally. Insulation resistance
Insulation resistance for instruments using AC power should not be less than 20MΩ.
Which gas detectors require regular calibration and verification?
Fixed-install, handheld, and portable gas detectors, gas alarm instruments, toxic and harmful gas detectors/alarm systems, flammable gas detectors with alarms, gas analyzers, etc.
What factors can affect the calibration cycle of a combustible gas detector?
Manufacturer Requirements: Different manufacturers may have varying inspection cycle requirements for flammable gas detectors. Therefore, when purchasing equipment, it's essential to carefully review the product manual to understand the specific requirements of the manufacturer.
Equipment Usage Environment: The operating environment of the alarm directly affects its inspection cycle. If the equipment is frequently exposed to high concentrations of harmful gases, high or low humidity, or high or low temperatures for extended periods, its inspection cycle may be shortened accordingly.
Equipment Maintenance Status: The condition of equipment maintenance is also a significant factor affecting the inspection cycle. If the equipment faces issues like battery depletion or sensor aging over a long period, the inspection cycle will accordingly be shortened.
Calibration Method
Appearance and structure
Inspect and feel in accordance with requirements of 4.1
Logos and Branding
Inspection to be conducted in accordance with requirement 4.2 by visual means.
Power-on inspection
Inspect visually and by touch, in accordance with 4.3 requirements, to check the 5.3.4 alarm function and alarm action values.
Introduce a standard gas substance with a concentration exceeding the alarm set point, causing the instrument to trigger an alarm. Observe whether the instrument's audible and visual alarms function properly, and record the instrument's reading at the time of the alarm. Repeat the measurement three times, and the arithmetic mean of the three readings is the instrument's alarm trigger value.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarm systems, and toxic/hazardous gas detection instruments for regular inspections, metrological measurements, and calibration services. We are qualified to address fire and explosion prevention issues for our clients.