Machine Model: LDQ-2 Touch Screen Controlled Leakage Current Tracking Tester 1. Leak Tracking Tester Application Range
Measure the relative resistance to electrical tracking of solid electrical insulating materials when exposed to impure water under the action of an electric field at voltages up to a maximum of 600V. When voltage is applied between electrodes placed on the surface of the material and the electrolyte is dripped between the electrodes at specified time intervals, electrical tracking may occur in this test condition. The number of liquid drops required to cause material failure increases with the decrease in applied voltage, and no electrical tracking occurs below a certain limit. When the material does not exhibit electrical tracking even at the highest test voltage, there can be varying degrees of corrosion, and the corrosion depth can measure that certain materials can ignite during the test.
II. Functional Features of Leakage Tracking Tester
Programmable Logic Controller Touch Screen + PLC Control
2. Voltage, start draining, switching, current, stop, exhaust ventilation, and lighting are all controlled via the touch screen. Test data can be stored within the touch screen. Upon completion of the test, the save interface will automatically pop up. Simply enter a save name to save the data. All saved data can be queried in the historical data section.
3. Voltage, Dropping Time, Current, Dropping Frequency, Equipment Door Status, Maximum Current, Real-time Display of Experimental Results
4. 7-inch full-color touch screen
5. Equipped with a top-mounted exhaust and anti-backflow device, allowing for the expulsion of smoke produced by combustion upon completion of the test.
6. Selectable test power range
7. Dropper frequency and time can both be set
8. Automatically Determine Test Qualification
9. Overcurrent Protection
10. Safety Protection: Automatic prompt of door status; no voltage in the testing area when the door is open.
11. Peak Current Value Alert III. Main Technical Parameters of Leakage Tracking Tester
1. Electrode Materials: Test Electrode - Platinum, Electrode Rod - Copper (Platinum Purity Meets Standard Requirement of 99%)
2. Electrode Dimensions: (2mm ± 0.1mm) x (5mm ± 0.1mm) x (40mm ± 5mm)
3. Platinum electrode, 12mm, 30°±2° bevel
4. Electrode Spacing: 4.0mm ± 0.01mm, Angle 60° ± 5°
5. Electrode Pressure: 1.00N ± 0.01N
6. Test Solution Resistance: A Solution - 0.1% NH4Cl, 3.95±0.05 Ωm; B Solution - 1.98±0.05 Ωm
7. Drip Device: No adjustment required during testing, simple operation
8. Droplet Height: 35mm ± 5mm (Adjustable)
9. Droplet Time: 30s ± 0.01s (Better than Standard) (Can be preset and adjusted)
50 drops time 24.5min ± 2min
Drop Count: 1 to 9999 (presettable) (adjustable drop size, with a capacity of 44-55 drops equal to 1 cm³)
10. Test Wind Speed: 0.2 m/s (New Standard)
11. Test Voltage: 100V ~ 600V (25V graduation, adjustable)
Power Drop: 8% at 1.0A ± 0.1A
13. Stain Judgment: 0.50A ±10%, 2.00s ±10%
14. Voltage adjustable between 100-600V (48-60Hz), voltage drop not exceeding 10% when short-circuit current is 1.0±0.0001A (better than standard 0.1A).
15. Exhaust Vent: Ø100mm
16. Air Environment: Temperature 0~40℃; Relative Humidity: ≤80%; No significant vibration and corrosive gases.
17. In the test circuit, if the short-circuit current exceeds 0.5A and remains for 2 seconds, the relay will activate, cut off the current, and indicate that the test sample is不合格.
18. Test Power: 220V 0.6kVA 50-60Hz
19. Dimensions: Width 1160mm x Depth 600mm x Height 1310mm (including feet)
Test Area Volume: >0.5 cubic meters, dimensions approximately 900mm x 590mm x 1050mm, background color black, background illuminance ≤ 20 Lux