Thermo Orion OXY-DO dissolved oxygen electrode 083005MD:
Application of 083005MD: Laboratory and Field Measurement
Cable Length: 1.5 m
Matching Host Machine: Star Series
Sample low flow rate: 10 cm/sec
Sample Temperature: 0 - 50℃
Electrolyte usage: 180 days
Response Time: Achieve 90% within 10 seconds, 95% within 15 seconds, and 99% within 1 minute.
Attachment Model: Film Sleeve - 080515, Electrolyte - 080514, Maintenance Kit - 080513
Oxygen Saturation Measurement Basics
· When measuring dissolved oxygen in the lab, the solution being tested must be stirred at a constant speed; for outdoor measurements, the water body must have a certain flow rate to ensure the rate of dissolved oxygen passing through the electrode membrane.
· The solution must not generate an electromagnetic field that interferes with measurement signals during stirring or flow, to prevent instability in measurement data.
· Oxygen sensors require a certain polarization time before calibration and measurement.
· When maintaining the oxygen electrode, it is essential to first turn off the main power of the unit, then remove the electrode for maintenance.
Order Number Description
080513 Dissolved Oxygen Maintenance Kit (1 bottle of electrolyte, 1 sheet of polishing paper, 2 membrane sleeves)
083005MD, 083010MD, 083025MD, 083060MD, 086030MD, 083005A, 083010A,083025A, 083060A,083005D, 086020A, 083150A, 083010F, 081010F, 081030F
080515 - Case 1 piece
083005MD, 083010MD, 083025MD, 083060MD, 086030MD, 083005A, 083010A,083025A, 083060A, 083005D, 086020A, 083150A, 083010F, 081010F, 081030F
081003 - 3 Membrane Sleeves
080510MD, 081010MD, 081010
Electrolyte Solution
083005MD, 083010MD, 083025MD, 083060MD, 086030MD, 083005A, 083010A,083025A, 083060A, 083005D, 086020A, 083150A, 083010F, 081010F, 081030F,080510MD, 081010MD, 081010