Thermo Orion OxyGuard Dissolved Oxygen Electrode 083005MD:
083005MD Application: Laboratory and Field Measurement
Cable Length: 1.5 m
Matching Host Machine: Star Series
Sample low flow rate: 10 cm/second
Sample Temperature: 0 - 50℃
During electrolyte use: 180 days
Response Time: 90% achieved within 10 seconds, 95% within 15 seconds, and 99% within 1 minute.
Attachment Models: Film Sleeve - 080515, Electrolyte - 080514, Maintenance Kit - 080513
Oxygen Dissolved Measurement Basics
· During laboratory measurement of dissolved oxygen, the solution to be tested must be stirred at a constant speed; when measuring outdoors, the water body must have a certain flow rate to ensure the rate of dissolved oxygen passing through the electrode membrane.
· The solution must not generate an electromagnetic field that interferes with measurement signals during stirring or flow, to prevent instability in measurement data.
· Oxygen Saturation Electrode requires a certain polarization time before calibration and measurement
· When maintaining and repairing the dissolved oxygen electrode, the main power must first be turned off, then remove the electrode before proceeding with maintenance.
Order Number Description
080513 Dissolved Oxygen Maintenance Kit (1 bottle of electrolyte, 1 sheet of polishing paper, 2 membrane sleeves)
083005MD, 083010MD, 083025MD, 083060MD, 086030MD, 083005A, 083010A,083025A, 083060A,083005D, 086020A, 083150A, 083010F, 081010F, 081030F
080515 - Protective Covering 1 Piece
083005MD, 083010MD, 083025MD, 083060MD, 086030MD, 083005A, 083010A,083025A, 083060A, 083005D, 086020A, 083150A, 083010F, 081010F, 081030F
081003 - 3 Membrane Sleeves
080510MD, 081010MD, 081010
Electrolyte
083005MD, 083010MD, 083025MD, 083060MD, 086030MD, 083005A, 083010A,083025A, 083060A, 083005D, 086020A, 083150A, 083010F, 081010F, 081030F,080510MD, 081010MD, 081010