Model LHS-010VA
Section II: Equipment Technical Specifications
Item | Performance Parameters | |
TechArtIndicatePlease provide the Chinese content you would like translated into American English. | Studio size | 1000x1000X1000 mm |
Outer Carton Dimensions | One side less than 1.4 meters | |
Temperature Range | Temperature Range: +15°C to +50°C | |
Humidity Range | Humidity Range: 30% to 80% RH | |
Instrument resolution | Temperature: 0.1°C | |
Humidity: 0.1%RH | ||
Variance | Temperature: ≤0.5℃ | |
Uniformity | Temperature: ≤1℃ | |
Temperature Deviation | Temperature: (-0.5°C, 0.5°C) | |
Humidity Deviation | Humidity: (-2%RH, 3%RH) | |
Studio Wind Speed | The center of the box has a wind speed of 0.1 to 1 m/s. (Adjustable wind speed) | |
Air Exchange Rate Range | 0.2 to 2 times per hour | |
Flow adjustment range must be greater than 0.2 to 2 m³/h, i.e., 0.2 is not the min value, 2 is not the flow min value, and the accuracy error of the extreme values is relatively large. | ||
Flowmeter accuracy | Glass转子流量计, accuracy not less than 2.5 grades | |
Sealability | Method 1: At overpressure of 1000pa, the leakage is less than 10-3×1m3/min Method Two: Gas flow difference between import and export is less than 1% | |
Stable time | Temperature stabilization time less than 40 mins; humidity stabilization time less than 50 mins, equipment can run continuously for at least 30 days. | |
Baseline Indicator | Formaldehyde background concentration: ≤0.006 mg/m³ Single VOC background concentration: ≤0.002 mg/m³ TVOC background concentration: ≤0.02 mg/m³ | |
Noise | No resonance sound on the box door equipment door at or below 65 decibels. | |
Other | Ensure no liquid or solid water is present inside the working chamber. | |
BoxStructureFeatures | Structural method | The inner lining corners feature large chamfer design, and the work chamber is fully welded for enhanced sealing performance. |
Composition of Materials | Outer Material: 1.5MM cold-rolled steel plate with electrostatic powder coating, color as shown in the picture. Inner Wall Material: 1.2mm, Mirror Finish SUS304 Stainless Steel Surface finish: III D, all internal interfaces are polished Insulating Materials: Natural mineral wool, thickness not less than 80mm, excellent thermal insulation properties. Sealing Materials: Fluorinated rubber, etc., VOC-free organic compounds Pipe connections: Materials for the intake section in contact with the interior compartment must not be organic. All materials inside the box do not adsorb or release formaldehyde and VOCs. | |
Recirculating fan unit | Choose all-metal blade axial flow fans with oil-free bearings; install 2 fan exhausts at the bottom back side (the wind plate is lower, with the diameter edge 1500MM from the bottom), exhaust through a porous plate at the top, the recirculating air is easy to disassemble and clean; the sharp section is rounded to prevent injury and is easy to clean. | |
Studio | The work cabin is integrally molded with seamless welding, the weld surfaces are polished for a smooth finish, ensuring no blind spots or gaps. | |
Hatch door | Hinge door, multi-point clamping, adjustable gap. The design features a double-door opening, ensuring no air exchange between the inner chamber and the outer environment. There is a heating element between the door and the glass to prevent condensation from forming when the internal temperature of the box is higher than the external temperature. | |
Observation window | Inner door centered position, with 3-layer hollow glass, featuring a heating element and anti-condensation function. | |
Sample hole | On the left side of the equipment, there are 3 air sample collection points inside the cabin. The sampling ports are made of finely processed brass material, and the sampling pipes are connected to the inner stainless steel panel through stainless steel pipes, without touching the jacket layer in between, thereby not being affected by the air in the jacket layer. Charcoal or TENAX sampling tubes can be directly inserted into the test cabin for sampling. To avoid external air affecting the sampling port, a 6x9 rubber tube can be matched and connected, or a sampling tube with a diameter of 0.8mm can be directly inserted, and the sampling port can be sealed with a spiral seal. | |
Luminous lamp | Utilizing LED light strips, designed on the observation cabin exterior inner door, for easy observation of samples inside. | |
Air Conditioning | Supply of gas | Electromagnetic compression air pump, with an independent power plug design |
Temperature Control | The temperature control method of the air sleeve, which is an air sleeve between the inner and outer casings, is equipped with cooling coils and heating pipes. | |
Humidity Control | Humidity control is achieved through a dry-wet mixing method, blending air dried by an oil-free air compressor with moist air to reach the set humidity level. | |
Intake Volume Control | Flow meter needle valve, intake through porous tube in the working chamber for even release | |
Air inlet purification | The air pump is followed by a series connection (activated carbon), and a molecular sieve filtration unit. | |
Refrigeration
| Refrigeration Compressor | 2 fully enclosed compressors (imported from France, Tai Kang) |
Heat exchange method | Finned evaporators, installed within the recirculating air duct. | |
Condensation Method | Air-cooled | |
Refrigerant | Environmentally Friendly Refrigerant R404a | |
Humidity Control | Humidity and temperature sensor, temperature 0-7 | |
Sensor | Temperature Control Sensor | Sheathed Platinum Resistance Thermometer (Imported) |
Temperature Transmitter | PT100, humidity sensor (imported) | |
EquipmentControl | Controller | 7-inch Color LCD Touchscreen Temperature and Humidity Controller Control Mode: PID (Proportional-Integral-Derivative) automatic calculation, PID control is simple and easy to understand. |
Human-Machine Dialogue Interface | Colorful LCD (7-inch) Display, Touch Key Settings | |
Instruction Precision | Temperature: 0.1°C, Humidity: 0.1%RH, Time: 1s | |
Operation | The product features a resistive high-precision touch screen for operation, allowing direct use of a USB mouse or stylus to click on the screen for function selection or parameter setting. | |
Operating Mode | Automatic operation, manual operation. | |
Display Items | Temperature, Measured Temperature, Set Humidity, Measured Humidity, Total Operating Time, Segment Operating Time, Inflow Rate, including historical data query, real-time curve, and historical data playback, fault records | |
Established Project | Work cabin temperature, work cabin humidity, start-up time | |
Record feature | Work cabin temperature, humidity, time; equipped with USB port, supports USB flash drive storage and backup. | |
Data correction | Work cabin temperature and humidity are adjustable. | |
Others | Feature with permission level settings to prevent user errors | |
SafetyEquipment | Work cabin | Over-temperature Alarm Protection |
Refrigeration Compressor | Fan Overpressure Protection, Overload Protection | |
Grounding Protection | Safety grounding protection for structural components, insulation resistance of electrical components and equipment enclosures not less than 0.5M ohms | |
Heater | Heater Short Circuit Protection | |
Humidifier | Humidifier Low Water Protection | |
Condenser | Cooling Condenser Low Water Protection | |
Other: The test chamber is designed for water wash-down, with easy-to-remove air duct baffles. Hatch coaming and piping associated with intake and exhaust ducts, as well as water towers and internal materials, are all non-organic to prevent contamination. | ||
Section 3: Equipment Appearance Images

IV. Equipment Design Advantages
Design Advantages1:
The variable frequency speed controller for wind speed adjustment is located at the rear of the housing, designed to be recessed for ease of access and less prone to accidental contact. |
2. Place the fans at the very back, two fans, positioned near the bottom. |
3. The vent hole diameter of the airfoil is reduced to approximately 2-3mm. |
4. The vent holes at the bottom are modified to a louver-like shape, making the internal compressor invisible from the outside, while the ventilation holes on the other three sides are very small. |
5. Electrical loads are isolated from each other. |
6. The case is well-filled with material, especially near the door, ensuring minimal noise when the door opens and closes. |
Design Advantages2:
1. Door handle assembly machinery must have excellent performance. |
2. Sampling hole3Individual units feature flanged connections at the inner and outer diameter junctions. |
3. Foot Wheel Support Triangular Board, Thickened5mm。 |
4. The internal impeller vent holes are not protruding externally; they are embedded within the wind ring. |
5. The box door features a double door swing-style opening. |
6. Electrical cabinet doors feature top and bottom pins, no screws required; reinforced ribs are added to the door. |
7. The internal fan positioning is accurately adjusted. |
8. The equipment features a computer white color, offering bright and vibrant hues. |