- AllProduct Category
-
HG30 Series Multi-Function Calibrator
DO30 Series Multi-Function Meters Calibration Device
XF30 Series Multi-Function Calibrator
HGZX Series Standard Resistor Box
HGGC Series Process Verifier
Megohmmeter Calibration Equipment
Three-Phase Multi-Function Standard Power Supply
HGSZ Series Distortion Meters
Fluke Precision Digital Multimeter
Huaguang Measuring Instruments Series
HG30 Series Multi-functional Calibrator


详情描述
I. Purpose and Features:
XF30-I Type DCMulti-functional Calibration InstrumentSuitable for calibration and verification of various DC current, voltage, and power meters below 0.5 class. Also serves as a high-stability test power supply, used in conjunction with high-grade standard meters to calibrate DC current, voltage, and power meters above 0.5 class. With optional standard test coils, it can verify DC clamp meters below 1 class.
II. Main Functions:
1: Five-Digit LED Digital Display for Output Quantity
2: Direct current output range: 0-30A, divided into 11 steps, with a resolution of 0.01uA
3: DC voltage output range: 0~1000V, divided into 11 steps, resolution of 0.01mV
4: Power measurement range: 0~30kW, resolution: 0.0001W
5: Overload protection with automatic shutdown and manual reset
Each voltage and current level can withstand a 10% overload.
Section 3: Technical Performance
Stability: Voltage < FullRange0.01% / 3 minutes
Current < FullRange0.02% / 3 minutes
2: Direct Current Ripple Factor: <0.1%
3: Output voltage, current, power range, and accuracy (at 23℃±2℃, output value greater than 10%)Range)
Export Items | Output Range | Load Limit | Accuracy |
Direct Flow Voltage | 0~200mV~1V~2V~5V | 100mA | Voltage: ±(0.03% of reading) +0.02% Full Scale (FS) Current: ±(0.03% of reading) +0.02% Full Scale (FS) Power: ±(0.07% of reading) +0.03% Fullness (±0.03%) |
0~10V~20V~50V | 100mA | ||
0~150V | 80mA | ||
0~300V~600V | 60mA | ||
0~1000V | 60mA | ||
Direct Flow Electric Current | 0~200uA~2mA~5mA~20mA~ | 10V | |
50mA~200mA~500mA~ 2A~ 5A | |||
0~ 10A~ 30A | 1.5V | ||
Note: For every 10℃ increase in working environment temperature above 23℃, the additional error is less than the basic error of the range. | |||
4: Power Consumption: AC voltage 220V ± 10%, frequency 50Hz ± 1Hz; power consumption < 200VA
5: Working Environment: Temperature of the working environment should be between 5℃ and 35℃, with relative humidity not exceeding 80%.
6: Business Hours: Continuous
7: Dimensions: 139×480×420mm3
Weight: Approximately 16kg