- AllProduct Category
-
HG30 Series Multi-Function Calibrator
DO30 Series Multi-Function Meters Calibration Device
XF30 Series Multi-Function Calibrator
HGZX Series Standard Resistor Box
HGGC Series Process Verifier
Megohmmeter Calibration Equipment
Three-Phase Multi-Function Standard Power Supply
HGSZ Series Distortion Meters
Fluke Precision Digital Multimeter
Huaguang Measuring Instruments Series
HG30 Series Multi-functional Calibrator


详情描述
DC Voltage Technical Specifications(20℃±2℃):
Range | Stability1min) | Accuracy ± (Read value)%+Full Range%) | Wavy pattern Coefficient | Adjust Scope | Load Capacity High value |
75.000mV* | 0.01%RG | ±(0.02%RD+0.01%RG) | <1% | 110%RG | 100mA |
1.0000V | |||||
3.0000V | |||||
10.000V | |||||
30.000V | |||||
100.00V | |||||
300.00V | |||||
1000.0V | 20mA |
* Accuracy below 10mV: ± (0.03%RD+0.02%RG)
DC Current Technical Specifications(20℃±2℃):
Range | Stability1min) | Accuracy ± (Reading Value)%+Full scale%) | Adjust Scope | Adjust Fineness | Output Load Capacity | Wavy texture Coefficient |
0.1000mA | 0.01%RG | ±(0.03%RD+0.01%RG) | 110%RG | 0.01%RG | ≤10V (Less than0.5A) ≤1.2V (Greater than0.5A) | <1% |
0.5000mA | ±(0.02%RD+0.01%RG) | |||||
1.0000mA | ||||||
10.000mA | ||||||
100.00mA | ||||||
1.0000A | ||||||
5.0000A | ||||||
10.000A | ||||||
20.000A | 0.02%RG | ±(0.03%RD+0.02%RG) |
Direct Current Power Technical Specifications(20℃±2℃):
Accuracy | Stability1min) |
± 0.05%RG | 0.01%RG |
Direct Current Power Range: 1W ~20kW。
AC Voltage Technical Specifications(20℃±2℃):
Range | Load Capability | Adjust Fineness | Accuracy ± (reading value)%+Full Range%) | Distortion | |
45Hz-400Hz | 400Hz-1000Hz | ||||
200 mV | 800 mA | 50ppm | 0.03%RD+0.02%RG | <0.2% | |
2 V | 0.02%RD+0.01%RG | 0.03%RD+0.02%RG | |||
20 V | |||||
200 V | 500 mA | ||||
1000 V | 200 mA | 0.03%RD+0.02%RG | |||
Technical Specifications for Alternating Current Communication(20℃±2℃):
Range | Load Capability | Adjust Fineness | Accuracy ± (Read value)%+Full Range%) | Distortion | ||
45Hz-400Hz | 400Hz-1000Hz | |||||
2mA | 30V | 50ppm | 0.03%RD+0.02%RG | <0.2% | ||
20 mA | ||||||
200 mA | 50V | 0.02%RD+0.01%RG | 0.03%RD+0.02%RG | |||
2A | ||||||
10A | 20V | |||||
20A | 0.03%RD+0.02%RG | |||||
* 100mAThe following quantities can be measured using dedicated shielding test lines, which can enhance0.005%Accuracy.
Communication Power Technical Specifications(20℃±2℃):
Scope | Accuracy | Stability1min) |
0W~20kW | PF=±0.8~±1Please provide the Chinese content to be translated.(50Hz),<0.05%FS PF=±0.8~±0.5Time50Hz),<0.1%FS | 0.01%RG |
Communication Power Range: 1W ~20kW
Power factor range:+1~-1Accuracy: <0.05%RG
Phase Output Technical Specifications:
Adjustable Range | Adjust fineness | Accuracy |
0~359.999° | 0.001° | 0.05° |
Frequency output technical specifications, as shown in the table below:
Adjustment Range | Accuracy |
45Hz~400Hz | ±0.05 Hz |
400Hz~1kHz | ±0.2 Hz |
Resistor Technical Specifications(20℃±2℃):
Range | Resolution | Short-term stability (%RG/min) | Accuracy within one year ± (Reading Value)%+Full Scale%) |
20Ω | 1mΩ | 0.02 | 0.05%RD+50mΩ |
200Ω | 10mΩ | 0.01 | 0.03%RD+0.02%RG |
2kΩ | 0.1Ω | ||
20kΩ | 1Ω | ||
200kΩ | 10Ω | ||
2MΩ | 0.1kΩ | ||
20MΩ | 1kΩ | 0.05 | 0.12%RD+0.08%RG |
200MΩ* | 10kΩ | 0.1 | 0.6%RD+0.4%RG |
Please use a dedicated shielded test cable for resistance measurement.
* To ensure the stability of resistance measurement values,1kPlease use the four-line measurement method.
Measurements above 100MΩ are possible but not assessed for accuracy.
Other Technical Specifications:
Please provide the Chinese content you would like translated into American English. Quantity:≤ 13kg(Excluding attachments and packaging boxes)
Body Accumulated:400mm×370mm×182mm
Operating Power Supply: AC220V±10%
Total Power Consumption:≤ 150W
Work Environment: Temperature-20~40°C; Relative Humidity≤85%。
询价单