详情描述

I. Application and Features:
XF30-IIa Model ACMulti-functional Calibration Instrument(Improved) Suitable for calibration and verification of various AC current, voltage, frequency, phase, power, and power factor meters below 0.5 class. Also serves as a high-stability test power source, compatible with high-grade standard meters, for calibration of AC current, voltage, and power meters above 0.5 class. With standard test coils (optional), it can verify AC clamp-on current meters below 1 class (0~1000A).
II. Main Functions:
1: Five-Digit LED Digital Display Output Quantity
2: The交流current output range is 0~100A, divided into 11 steps, with a resolution of 0.1uA.
3: Output voltage range for communication: 0~1000V, divided into 11 steps, resolution of 0.1mV
4: Power measurement range: 0~30kW, resolution: 0.0001W
Frequency measurement range: 40~70Hz
Power factor range: 0~±1; Phase range: 0~360°
7: Overload protection with automatic shutdown and manual reset
Each voltage and current level can withstand an overload of 10%.
9: Output can be selected as "Local Oscillator/Sync", with the local oscillator frequency ranging from 40 to 70 Hz, and the sync frequency is synchronized with the municipal electricity.
10: Displayed in actual value or percentage form
III. Technical Performance:
Stability: Voltage, current < full loadRange0.01% / 3 minutes
Power < FullRange0.02% / 3 minutes
Additional Frequency Impact 0.005%
2: Distortion: <0.5%
3: Output voltage, current, power range, and accuracy (at 23℃±2℃, output value greater than 10%)Range)


Export Items

Output Scope

Load Limit

Accuracy

Communication
Voltage

0~1V~2V~5V~10V

800mA

Voltage: ±(0.03% of reading)
+0.02% Full Scale)
Current: ±(0.03% of reading)
+0.02% Fullness)
Power: ±(0.07% of reading)
+0.03% Fullness)
Angle: ±0.2°
Power Factor: ± 0.001

0~20V~50V

500mA

0~75V~150V

200mA

0~300V~600V

80mA

0~1000V

40mA

Communication
Current

0~10~20~50~200mA

3V~40V

0~1~2~5~10~ 25A

2V~20V

0~50~ 100A

1V

Note: 1. For working environment temperatures exceeding 23℃ ± 2℃, the additional error is less than the basic error of this range for every 10℃ change.
2. No assessment for phase angle and power factor is required for currents of 200mA or less. The power factor assessment range is ±0.5 to ±1.


4: Output Frequency Accuracy: ±0.01 Hz
5: Output phase accuracy: ±0.2° (after zeroing)
Power Consumption: AC voltage 220V±10%, frequency 50Hz±1Hz; power consumption < 300VA
7: Working Environment: Temperature in the working area should be between 5℃ and 35℃, with relative humidity not exceeding 80%.
8: Office Hours: Continuous
9: Dimensions: 179×480×420mm3
Weight: Approximately 20kg

询价单

主题 *
内容 *
Your company's名
Contact person *
ContactPhone *
电子Email
验证码  
 点击确定代表您同意《服务条款》《隐私政策》