- AllProduct Category
-
HG30 Series Multi-Function Calibrator
DO30 Series Multi-Function Meters Calibration Device
XF30 Series Multi-Function Calibrator
HGZX Series Standard Resistor Box
HGGC Series Process Verifier
Megohmmeter Calibration Equipment
Three-Phase Multi-Function Standard Power Supply
HGSZ Series Distortion Meters
Fluke Precision Digital Multimeter
Huaguang Measuring Instruments Series
HG30 Series Multi-functional Calibrator


详情描述
Portable Multi-Function Calibration Instrument | Portable Multimeter Calibration Instrument | Portable Multifunction Meter Calibration Instrument | Portable Multimeter Verification Instrument | Portable Multimeter Calibration Instrument | Portable Multimeter Verification Instrument | Portable Calibration Instrument | Portable Verification Instrument
Purpose and Features:
The HG5610A Multi-functional Calibration Instrument (Tri-Function Meter Calibrator) is suitable for verifying various current and voltage meters below 0.2 class. It can also serve as a high-stability test power supply, used in conjunction with high-grade standard meters to calibrate current and voltage meters above 0.2 class.
2. Main Functions:
2.1, 5 1/2-digit LED digital display for output quantity.
2.2 Output range of AC and DC current: 0~50uA~5mA~50mA~500mA~2A~10A
2.3 Output voltage range for AC and DC: 0~500mV~5V~50V~250V~500V~1000V
2.4. The available frequencies for selection are three: 50Hz, 60Hz, 400Hz
2.5 Output overload can be automatically protected and manually reset.
2.6. Each voltage and current level can withstand an overload of 10%.
2.7 Clamp Meter Measurement: Equipped with our standard coil, capable of measuring AC 0-750A (some models can measure up to over 1000A); DC 0-1000A with an error of ±0.3%.
3. Technical Performance
3.1 Stability: DC < 0.01% of full scale / 2 minutes
AC < 0.03% FS / 2 minutes
3.2 Distortion in Communication: <0.5%
3.3 DC Ripple Factor: <0.1%
3.4 Output Frequency & Accuracy: 50Hz 60Hz 400Hz <1%
3.5 Output Voltage and Current Range and Accuracy (at 23℃±2℃, output value greater than 10% of the range)
Export Items | Output Range | Rated Output | Peak Output | Accuracy | |
AC/DC
Voltage | 0~500mV | 10mA | 30mA | DC: ±(0.02% of reading + 0.03% of full scale) AC: ± (0.03% of reading + 0.05% of full scale) | |
0~5V | 20mA | 60mA | |||
0~50V | 40mA | 100mA | |||
0~250V | 20mA | 60mA | |||
0~500V | 10mA | 30mA | |||
AC/DC Electric Current | 0~50uA~5mA~50mA | DC:3V AC:36V | DC: ±(0.02% of reading + 0.03% of full scale) AC: ±(0.03% of reading + 0.05% of full scale) Below 2mA, accuracy is not assessed. | ||
0~500mA | DC:3V AC:12V | ||||
0~2A~10A | 2V | ||||
Standard Resistance | 10、50、100、 500(Ω) 1K、50K、100K、500K、 1M、10M、20M、 |
0.25W |
±0.3%+20mΩ |
3.6 Power Consumption: AC power voltage 220V±10%, frequency 50Hz±1Hz; power consumption < 180VA
3.7. Work Environment: Temperature 5°C to 35°C, relative humidity not exceeding 80%
3.8 Work Hours: Continuous
3.9 Dimensions: 135×400×360mm
3.10 Weight: Approximately 12kg
询价单