详情描述

Far-Infrared Carbon Dioxide Incubator HH.CHP-TW - Cell CultureThe daily maintenance includes the following points:

Regularly monitor the temperature and humidity inside the box to ensure the incubator operates normally.
Check that the box door is securely closed. If any gaps are found, immediate action should be taken to resolve the issue.
3. Ensure the exhaust fan is well-ventilated; regularly clean the fan blades for smooth airflow.
4. Regularly inspect power cables to prevent overloading.
5. Regularly clean the condenser of dust to maintain its efficient heat dissipation. Preferably use a damp cloth wrung out of water to wipe, to avoid water渗入 the interior of the machine case.
6. Always disconnect the power source when not in use and ensure that any water inside the box is thoroughly drained.
7. During maintenance, ensure personal safety and avoid operating equipment without a full understanding of it.


Far-Infrared Carbon Dioxide Incubator HH.CHP-TW - Cell CultureKey Features:

1. The CO2 Incubator features a streamlined curved design, with the outer shell made of cold-rolled steel plates and electrostatically powder-coated finish. The inner chamber is constructed from stainless steel material, and the semi-circular corner design facilitates easy cleaning.

2. Microcomputer temperature controller with minimal temperature fluctuations, equipped with an ultraviolet sterilization lamp inside the box for regular UV disinfection, preventing contamination during cell culture.

3. Double-door structure: After the outer door is opened, observe the laboratory conditions inside through a door made of high-strength tempered glass, with no impact on temperature and humidity.

4. The CO2 concentration sensor features a Finnish far-infrared probe, allowing for direct display of the box's CO2 levels and operates reliably.

5. An independently set door heating system effectively prevents condensation on the inner door glass.

6. The studio utilizes a water tray for natural evaporation humidification, with humidity displayed directly by the gauge.

7. The CO2 incubator is equipped with an independent over-temperature alarm system. It automatically shuts off upon exceeding the limit temperature, ensuring safe operation of experiments and preventing any accidents.

 

Far-Infrared Carbon Dioxide Incubator HH.CHP-TW Cell CultureTechnical Specifications

Model

HH.CHP-T

HH.CHP-01

HH.CHP-TW

HH.CHP-01W

Volume

80L

160L

80L

160L

Heating Method

Gas-tight suit

Water jacket

Temperature Control Range

Room temperature +5~65℃

Temperature Resolution

0.1℃

Constant Temperature Fluctuation

±0.2℃

CO2 Control Range

0~20%

CO2 Control Method

Import Infrared Microcomputer CO2 Concentration Probe Controller

Humidification Method

Natural Evaporation (Water Tray)

Humidity Range

Greater than 95%RH (+37℃ operating temperature)

Office Hours

1 to 9999 minutes or continuous

Power

400W

500W

850W

1250W

Operating Power Supply

AC 220V 50Hz

Studio size

400*400*500

500*500*650

400*400*500

500*500*650

Dimensions

550*610*785

650*710*935

550*610*775

650*710*925

 


询价单

主题 *
内容 *
Your company's名
Contact person *
ContactPhone *
电子Email
验证码  
 点击确定代表您同意《服务条款》《隐私政策》