Xiangtan XiangkeDRX-3Hotline Thermal Conductivity Tester, High-Temperature Thermal Conductivity Instrument for Refractory Materials
Xiangtan XiangkeDRX-3Thermal Conductivity Tester
I. Overview
DRX-3Thermal Conductivity Test Instruments utilize the transient hot wire method, including cross-line and parallel-line techniques, primarily measuring the thermal conductivity, thermal diffusivity (heat transfer coefficient), and specific heat capacity of solid and powdered materials at various temperatures. They feature high precision, high resolution, and high reproducibility. The instruments are computer-controlled for automatic measurements. Widely applicable in research institutions, quality inspection, and industrial departments for material thermal conductivity measurements.
This instrument is a product of our factory.DRX-2Revised development and design on the basis, enhancing measurement accuracy and reproducibility, with a more rational structure and easier usage. Model Classification:DRX-3APrimarily utilizes the cross-hotline method (parallel line method can also be used for measurement)5w/mkThe following materials);DRX-3BUtilizing the parallel line method, measure less than25w/mkMaterials.
Instrument Reference Standard:GB/T 5990-2021"Fire-resistant Materials - Test Methods for Thermal Conductivity, Specific Heat Capacity, and Thermal Diffusivity"(Hotline Law)》;GB/T 17106-1997"Thermal Conductivity Test Method for Refractory Materials (Parallel Line Method)"GB/T 11205-2009"Determination of Thermal Conductivity of Rubber (Transient Wire Method)"; GB/T10297-2015"Determination of Thermal Conductivity of Non-Metallic Solid Materials"(Hotline Law)Etc.
Section 2: Key Technical Specifications
1Measurement Range:DRX-3A(Cross-line Method) 0.015~2.0 W/(m·K)
DRX-3A(Parallel Line Method) 0.015~5.0 W/(m·K)
DRX-3B(Parallel Line Method) 0.015~25.0 W/(m·K)
2Temperature Range: Routine testing is conducted at room temperature.Available for selection according to customer requirements are various thermostatic devices and test hotlines, enabling testing at different temperatures. Additional charges apply, as follows:
-40~120 ℃With high and low temperature constant temperature box, equipped with copper-Cupronickel Hot Wire
Room Temperature1000 ℃(Needed with NiCr resistance wire) NiCr equipped-Nickel Silicon or Platinum-Platinum Rhodium Alloy Hot Wire
Room temperature1300 ℃Supplied with a platinum-silicon carbide rod resistance furnace-Platinum Rhodium Alloy Hot Wire
Room temperature ~1600 ℃Supplied with a platinum-lined silicon molybdenum rod furnace-Platinum Rhodium Hot Wire
3Measurement Accuracy:<±5.0 %
4Sample Size:2Piece - Not less than40mm×80mm×114mmThe cubes or rectangles that are mutually symmetrical2Block cross-sectional diameter not less than80mm, Length not less than114mmA cylinder formed by stacking half-cylinders; recommend standard brick dimensions230 mm×110mm×65 mmOr230 mm×110 mm×75 mm (Or custom-made to customer's specifications)
5Temperature Discrimination:0.001℃
6Hotline Power Supply:DRX-3ACurrent Discrimination0.01mAVoltage differentiation0.1mV
DRX-3BCurrent Discrimination0.1mAVoltage differentiation0.1mV
7Equipment Power Supply: Voltage220VFrequency50HzPower500W(Refer to the furnace temperature controller manual for power requirements of the electric furnace temperature controller)
Standard Configuration: IncludesDRX-3One test host, one software package, one set of communication interfaces and data cables, one desktop computer, and one set of ambient temperature test hotlines.Optional items include thermostatic devices and high-temperature heating lines, as per the contract configuration.