I. Product Application
The LT-601 computer-controlled rubber vulcanometer is suitable for the rubber processing industry to control rubber quality, rapid testing, and fundamental rubber research. It provides accurate data for optimizing rubber formula combinations, accurately measuring parameters such as the vulcanization time, vulcanization index, and torque values.
Section 2: Applicable Standards: GB/T16548, ASTM D5289, and other testing standards.
III. Performance Features
1. This instrument features a true sealed mold cavity, offering repeatability and test data comparable to Alpha, placing it at the forefront of the industry.
2. This instrument is developed on a large database development platform, featuring a temperature control unit that is directly controlled and processed by software, ensuring stable, fast, and accurate temperature control.
LT-601 ModelRubber curing testerThe control and measurement method utilizes a human-machine interface, featuring a large LCD screen display with user-friendly design and easy operation.
4. Power System: Excitation motor + amplitude adjuster. Swing Angle: ±0.5º.
5. Data Transmission Method: USB Transfer.
6. Molds are made of Cr6WV material, featuring high hardness and non-deformation characteristics.
IV.Rubber Vulcanization InstrumentKey Technical Parameters
1. Temperature Range: Room Temperature - 200℃
2. Temperature Accuracy: ±0.3℃
3. Temperature Control Accuracy: ±0.3℃
4. Temperature Resolution: 0.01℃
Torque Range: 0-10 N·m, 0-20 N·m
6. Torque Resolution: 0.001 N.m
7. Molding Frequency: 1.7Hz
8. Amplitude: ±0.5°
9. Environmental Temperature: 0-35℃ Relative Humidity: <80%
10. Vulcanization Time Setting Range: 2-300 min
11. Pressure Regulating: 0.5 MPa
12. Power Supply: AC 220V, 800W
13. Machine Dimensions: Approximately 540×570×1150mm
14. Machine Weight: Approx. 280 Kg