JZ-6000Rapid Plasticity Tester
Rapid Plasticity Tester for testing natural rubber and unsulfured plastics(Blended rubber)Rapid Plasticity Value(Plastic Initial ValueP0)Maintainability of plasticity(PRI)This instrument consists of the main unit and a blanking machine.(Cutting Blade)Precision aging test chamber, thickness gauge, and other components. Rapid plasticity valueP0By rapidly compressing the cylindrical specimen between two parallel blocks through the main machine1mmThe fixed thickness, the tested sample maintains under compression state.15sTo achieve temperature equilibrium with the parallel plates, then apply a sample100N±1NConstant pressure, and maintain15sAt the end of this stage, the measured test thickness is used as a measure of plasticity by observing the gauges.
Plasticity Retention RatePRIThrough high-precision aging chambers, the specimens are subjected to140°Temperature aging30minAfterward, measure the plasticity value.P30Derived through calculationPRIValue.PRIValue is a measure of the natural rubber's resistance to thermal oxidation; the higher the value, the better the thermal oxidation resistance.
JZ-6000Rapid moldingPerformance is stable. This machine uses electronic direct temperature control, with fast heating and high accuracy. Additionally, it directly loads standard weights, overcoming the drawbacks of instability and zero-point drift caused by the electromagnetic spring pressure of the load. Synchronized with the Wallace's updated generation products.。
II. Execution Standards
GB/T3510? ISO2007? GB/T3517? ISO2930
III. Technical Specifications
AMain Engine:
1Temperature Control:100±0.5℃
2Loading Value:100±1N
3Power Voltage:AC 220V±10%
4Top pressure plate:φ7.3±0.02mm? φ10±0.02mm? φ14±0.02mm? Effective Height3.2mm
5Down Pressure Plate:φ15mm
BAging Box
1Temperature Control:140±0.2℃
2Power Voltage:AC 220V±10%
3Timing Duration:30min±5s