JZ-6029 Jiangdu Vulcanometer Model
One,JiangduSulphur AnalyzerKey Technical Specifications
Temperature Control Range: 0~200℃
Heating Time: ≤10min
Mold temperature fluctuation: ≤±0.1℃%
Torque Range: 0-20 N.m
Swing Frequency: 1.7 Hz (100 rpm)
Swing Range: ±0.5° ±1°
Mold cavity structure: Monsanto style
Report Content: ML, MH, ts1, ts2, t10, t30, t50, t70, t90, Vulcanization Curve, Temperature Curve
Standard: ASTM D5289-95, ISO 6502:1991, GB/T 16584-1996
Power Voltage: AC 220V ±10%
Dimensions: (Length x Width x Height) 680mm x 650mm x 1100mm
Net Weight: 230 Kg
Two,JiangduSulfur AnalyzerInstallation Guide
Position the main unit vertically at the location of use, adjust the ground screws at the bottom of the main unit, and use a level to check the worktable's levelness (longitudinal and lateral levelness of 0.2/1000).
Electrical InstallationYangzhou Jingzhuo VulcanometerAdjustments to measurement time, range, temperature control, adhesive batch number, team number, database, precision calibration, data retrieval, and curve data printing can be easily made on the display screen with simple settings and a few mouse clicks. The "Measure" button can complete multiple actions like "clamping," "turning," and "measuring" for quick operation. After measurement, the software automatically adjusts the Y-axis scale from 0-100Mv to 0-200Mv based on the measurement value size, enhancing the curve's aesthetics. The human-machine interface is highly user-friendly, with a clear visual guide on how to operate. The graphics are vibrant but not eye-straining.
III.JiangduSulphur AnalyzerTesting Requirements
1、Yangzhou Jingzhao VulcanometerIf tested every 30 minutes, with an 8-hour working day, the system can store at least 15 years' worth of curves and data. Therefore, aside from using the system for the Mooney viscosity meter, please do not use the computer for other purposes to prevent operational errors or virus infections, which could lead to data and system corruption, resulting in unnecessary losses.
2. The names of enterprises and institutions, names of materials for each curve, material batch number, team number, operator's name, etc., printed on the drawings from Yangzhou Jingzhuo vulcanometer, can be modified during measurement. If users need to add any title names for explanation, they can set any form of title text.
IV.JiangduSulphur AnalyzerImportant Notes
JiangduSulphur AnalyzerPrior to use, check that the instrument is properly installed in a room temperature environment with low relative humidity. Ensure there are no corrosive media nearby, the power voltage fluctuations are minimal, and the main unit must be grounded.