Primary Technical Performance
1. Display Range: 1nSv ~ 100Sv/h
2. Energy Range: 0.025 MeV ~ 23 MeV, 252Cf
3. Measurement Range: Dose rate 10nSv/h to 10Sv/h
4. Measurement Error: Not exceeding ±15% (typical value)
5. Sensitivity: 1 CPS/(µSv/h), 252 Cf.
6. Strong anti-interference capability, gamma suppression rate: >5µSv/h.
7. Response Time: <0.5S
8. Precision within 1% or 2 digits.
9. Dependency Degree: ±15% in all directions
10. The main unit can be externally connected to various probes.
11. Integrated GM counter tube
12. Automatic range switching, allowing multiple dose/dose rate alarm thresholds to be set across the full range.
13. Measurement values calculated and displayed include: dose rate, cumulative dose, peak dose rate, and average dose rate, which can be switched to display at any time.
14. Relative error of displayed measurement results.
15. Low power consumption: ≤300mW, single battery-powered operation for up to 1000 hours.
16. The housing is highly wear and corrosion-resistant, with a protection rating exceeding IP67.
17. Temperature: -10℃ to 50℃
18. Humidity: 95% (at 30℃)
19. Power Source: High Capacity 12V DC Lithium Battery
20. Dimensions: 31×32×13 cm
21, Weight: 3.5 kg.