High-Frequency Infrared Carbon-Sulfur Analyzer (JS-HW2000A Model) Carbon-Sulfur Analyzer 1. Main Technical Specifications of the Equipment 1.1 Measurement Range (1g Sample): Sulfur: 0.0001%-2.0000% 1.2 Instrument Analysis Accuracy: Carbon conforms to ISO 9556 standard; Sulfur conforms to ISO 4935 standard 1.3 Analysis Time: 25-60 seconds adjustable, usually around 35 seconds 1.4 Small Reading: 0.1 ppm 1.5 Electronic Balance (Imported Electronic Balance): Reading Accuracy: 0.0001g 1.6 Working Environment: Indoor Temperature: 10-30℃; Relative Humidity: less than 90% 1.7 High-Frequency Furnace: Power = 2.5kVA; Frequency: 18MHz 1.8 Power Supply: Requires good grounding, AC 220V ±5%; Frequency 50Hz ±2% 1.9 Oxygen: Input Pressure 0.18MPa ±5%; Oxygen-Carrying Pressure 0.08MPa ±2% 1.10 Gas Pressure: Nitrogen or compressed air (purified); Input Pressure 0.25MPa ±5% 1.11 Gas Flow Rate: Top Oxygen Flow Rate 1.0-2.0L/min; Analysis Gas Flow Rate 3.0-4.5L/min 1.12 Desiccant: High-Efficiency Color-Change Desiccant 1.13 Filter: Quartz棉 2. Main Features of the Equipment The JS-HW2000A Model Infrared Carbon-Sulfur Analyzer is used in conjunction with a high-frequency induction furnace to rapidly and accurately determine the mass fractions of carbon and sulfur in steel, iron, alloys, mold sand, non-ferrous metals, cement, ore, coke, catalysts, and other materials. This equipment introduces advanced foreign technology and is a high-tech product integrating optics, machinery, electricity, computers, and analysis technology. It features a wide measurement range, strong anti-interference ability, comprehensive functions, easy operation, and accurate and reliable analysis results.