Biosafety cabinet Class 2, airflow pattern: 30% exhaust, 70% internal recirculation required.
The outer casing is made of cold-rolled steel sheet with electrostatic spraying; the operation area features a monolithic stainless steel structure on all three sides. The washable internal areas have R-radius corner design for ease of cleaning. A liquid collection trough is installed at the bottom to facilitate cleaning and prevent cross-contamination.
Standard with two ULPA filters, achieving 99.995% filtration efficiency for particles as small as 0.12μm; the filter membrane is made of frameless borosilicate glass fiber.
The control system features a high-brightness, high-resolution color LCD touchscreen display, which实时 shows parameters such as intake wind speed, descent wind speed, sterilization, lighting, etc., along with settings for password modification and program configuration to prevent accidental operations.
The studio features a negative pressure leak-proof design, surrounding the work area with negative pressure on all four sides, truly achieving zero leakage and ensuring the cleanliness of the operation zone.
The fan comes with voltage fluctuation compensation and resistance induction compensation, effectively extending the lifespan of the filter.
Utilizes ultra-micro pressure difference wind speed sensors for real-time wind speed detection.
Temperature Sensor: Capable of real-time detection and display of temperature.
Equipped with filter pre-failure, damage, and blockage alarm technology, it promptly reminds of filter replacement.
Equipped with a height alarm, it alerts users with sound and light when the sliding glass door is below or above the specified height limit.
Features预约sterilization and sterilization timing function; records the cumulative operating time of the sterilization lamp and the filter. Includes a reminder to replace the ultraviolet lamp after more than 2000 hours of cumulative operating time (operating time can be set manually).
Integrated front window sliding door, freely positionable and able to be fully closed for sterilization. The front glass door is made of not less than 5mm thick safety tempered glass.
Front window and sterilization lamp interlock design: When the front window is opened, the sterilization lamp turns off, ensuring the safety of the experimenter and the experimental samples.
Parameters with memory retention and power failure, system crash state data protection, and power-on recovery function.
The front intake port features a design with unobstructed return air, ensuring effective prevention of backflow during operation, enhancing safety.
10-degree inclined design, more ergonomically designed for human dynamics, ensuring operator comfort.
In-cabinet Power Supply: Waterproof socket design for increased flexibility and convenience in operation.