1.Heating wire temperature: 500~1000℃±2℃ continuously adjustable 2.Burning time:0-999.9s±0.1sadjustable (The general choice is30s) 3.Ignition time:0-999.9s±0.1sAutomatic recording, manual pause 4.Extinguishing time:0-999.9s±0.1sAutomatic recording, manual pause 5.Thermocouple: Φ0.5mmImported armor(KType) thermocouple 6.Heating wire: Φ4 mmNickel chromium wire 7.Heat depth:7mm±0.5mm 8.Sample pressure:1N±0.2N 9.Test speed:14mm/s 。 10.Adopting a red line photoelectric stop switch,Make the limit more precise. 11.The thyristor controlled temperature control system ensures more accurate temperature regulation. 12.Time and temperature digital display, easy to observe and record 13.Experimental background: Black background. 14.Experimental process: Automatic control of the experimental program, independent ventilation 15.Studio volume:0.5 Cube, (stainless steel chassis can be customized according to customer requirements)0.75 Cube or 1 Cube) 16.Equipment external dimensions:1000mmWidth x650mmDeep x1300mmhigh
|
execution standardGB/T5169.10-2006 GB5169.11 GB4706