详情描述

I. Overview

The DWK-1500E explosion-proof constant current locatable medium flow flow gas sampler, independently researched and manufactured by Yancheng Dewang Instrument Equipment Co., Ltd. with patented technology, is a device capable of collecting gas or vapor samples from the atmosphere. It is suitable for individual toxic substance sampling of occupational hazards in workplaces and for dust sampling of metals and other particulate matter. It is also applicable to routine and emergency monitoring of environmental air, public health, and occupational health.

The Constant-Current Locatable Low-Flow Atmospheric Sampler is primarily composed of: the operating system, intelligent control system, air pump, flow measurement system, and positioning data upload module. It can continuously operate for over 15 hours on a single charge, convenient for sampling tasks. The instrument features real-time flow, sampling volume, sampling time, temperature, atmospheric pressure, and battery level displays. It is compact in structure, small in size, lightweight, durable, with strong load-bearing capacity, stable and accurate flow, and low noise. It is portable for easy use during sampling.

II. Application Fields

The DWK-1500E Constant-Current Positionable Low-Flow Sampler is an ideal instrument for short-term or long-term air sampling for disease prevention centers, environmental monitoring stations, colleges and universities, research institutions, and inspection agencies with environmental protection personnel.

Standard of Execution:

JJG956-2013规程 for Calibration of Atmospheric Samplers

GBZ/T300-2017 Specification for Sampling of Hazardous Substances in Workplace Air

GB50325-2020 Code for Indoor Environmental Pollution Control of Civil建筑工程

HJ 194-2017 "Technical Specification for Manual Monitoring of Ambient Air Quality"

III. Technical Specifications

Flow Range: (0-1500ml/min) (Flow adjustable) Resolution: 100ml/min

2. Constant Current Mode: Electronic Digital Constant Current , Supports various functions including data entry, mobile data collection, historical data download, data logic verification and audit, statistical analysis, and visual presentation. Equipped with a digital DTU positioning module, enabling real-time upload of sampling data, fault alarms, and display of sampling location coordinates and latitude/longitude. Flow is unaffected by voltage fluctuations and air resistance changes. Anti-  Absorbent Liquid Particle Suck back Equipment Our company-developed patented gas sampling device has a utility model patent. The primary function of this device is to prevent the absorption liquid from being siphoned back into the instrument. In case of back-siphoning, the device completely blocks the liquid from entering the instrument's interior, and it features an alert function. When the absorption liquid reaches the sampling device, a white circular sticker turns red upon contact with liquid, thereby alerting the user that back-siphoning has occurred, triggering the instrument's alarm and causing the fault light to flash. The device will automatically shut down after 30 seconds; at this point, simply pause the host. Open the water-stop device, clean it, replace the new water-stop membrane, and you can continue working. Therefore Significantly reduce Instrument Failure rate and convenient sampling work hours, no need for factory repair.

3. Flow Error: ≤±2.5%

4. Traffic Stability: ≤3%

5. Traffic Repetition: ≤2%

6. Timing Error: ≤±0.01%

7. Timing Range: 1min to 99h 59min

8. Operating Temperature: -10 to 50℃

9. Relative Humidity: ≤95%RH

10. Battery Life: Not less than 15 hours

11. Load Capacity: ≥35,000 Pa

12. Sampling Mode: ① Recurring Sampling (99 cycles), with options for recurring volume or time sampling. Set the sampling time, interval time, and number of cycles.

② Manual sampling, after setting the sampling flow, you can directly start sampling, with a running timer, and manually stop.

③ Schedule sampling, set the sampling time, and the sampling will stop once the time is reached.

④ Constant-volume sampling (1000L), set the sampling volume to 1000 liters.

12. (1) Lock Function: The panel lock feature prevents changes to the settings during pump operation.

12 (2) Flow Fault Correction: Fault alarm function display activated when flow variation exceeds ±5% within the back pressure specification range.

After 30 seconds, the error cannot be corrected, the pump will stop operating.

12 (3) The extra-large LCD screen displays: flow rate, sampling time, sampling volume, back pressure, battery expected sampling time, date, time, operation mode, and battery remaining charge.

12(4) Humidity Range: Operating humidity: 5-95%, non-condensing; Storage humidity: 5-98%, non-condensing

12 (5) Real-time access to the sampler's working data and actual conditions is available on our provided cloud platform.

13. Main Unit Dimensions: 130*44*95mm (Length*Width*Height)

14. Operating Noise: <55 dB

15. Instrument Weight: 0.52 kg

16. Power Supply: DC 7.4V, 220VAC ±10%

17. Volume Calculation: Real-time Display

18. Data Storage & Download: 1000 sets of real-time queryable sampled data

19. Explosion-proof Certificate No.: CNV20.0234 Explosion-proof Mark: Ex ib IIB T4 Gb