Optical Acoustic Gas Analyzer CF4

产品Price Negotiable

最小起订Quantity:1 Tai 供货总Quantity: 100 Tai

用手机扫描二维码直达商品手机版
MobileView details
Location:
Beijing
有效期至:
长期有效
最后更新:
2023-04-10 17:17
Product Details

Product Information

The GASERA ONE CF4 analyzer combines ultra-sensitive cantilever beam enhanced photoacoustic detection technology with a quantum cascade laser source operating at the fundamental mid-infrared spectral absorption line of CF4. This combination offers sufficient sensitivity to reliably detect the smallest CF4 leaks without delay. It also boasts high stability, with calibration cycles ranging from a few months to several years, resulting in a low total cost of ownership.


Technical Specifications

1. Working Principle: Photoacoustic Infrared Spectroscopy

Patented ultra-sensitive optical microphone measurement of the micro-movement of the cantilever light source structure based on MEMS cantilever sensors and laser interferometers: Distributed Feedback Quantum Cascade Laser gas cell stabilized to 50℃ temperature.

2. Standard

Compliant with Low Voltage Directive 2014/35/EU, EMC Directive 2004/108/EC, and RoHS 2 Directive 2011/65/EU.

3. Environmental

Operating Conditions: Temperature Range: 0°C...49°C, Humidity: Non-condensing Pressure Range: 750 mb...1050 mb Dust/Waterproof: IP20 (IEC529)

Storage Conditions: Temperature Range: -20°C to +60°C

Temperature: 0-49° C

Humidity: Non-condensing, with a maximum relative humidity of 80% at 35°C, linearly decreasing to 35% at 49°C.

Pressure: 750...1050 mbar; Gas Flow: Approx. 1 liter/min; Particle Size <1μm

4. Measurement Specifications

Response Time: Varies depending on the User Configurable Integration Time (C.I.T.) and gas exchange program, ranging from approximately 10 seconds to several minutes, depending on the average time and sampling details.

Detection Limit: Below 0.1 ppb

Dynamic Range: Over 5 orders of magnitude (i.e., repeatability: below 1% of the measured values under calibrated concentration operating conditions)

Accuracy: Better than 5% under calibration concentration. Limited by the precision of the calibration gas.

Temperature Stability: Environmental temperature variations within the operating temperature range will not cause drift.

Pressure Stability: Pressure fluctuations within the range of the sample gas will not cause drift.

5. General Cases

19 3U (Unit) enclosure, designed for desktop and rackmount installations.

Dimensions: 48, 4" W x 13.9" H x 44" D (19.1" W x 5.5" H x 17.3" D)

Weight: Approximately 13 kg

Built-in computer with a 7" WSVGA display, data storage capacity for at least 1 year of continuous monitoring, shortest sampling interval with an internal gas volume of 30 ml.

Input Voltage: 90...264Vac, 47...63Hz

Input Power: 75W max.

Interface: Ethernet, USB, and optional RS485, RS323, current message, voltage message, MODBUS, and AK protocol. The GASERA CF4 can be remotely operated via smartphones, tablets, laptops, or other GAS ERAONE devices.


Advantages

1. Accurate, reliable, low-maintenance control over cleanroom contamination with rapid response time, capturing rapid spikes, and an independently operated gas system内置

2. Trading Unit

3. No consumables or wet chemicals required

4. Offer short-path single-point calibration dynamic range for industry ling

5. Direct Absorption Measurement, No Drift Operation

6. Extended recalibration period (up to several years), highly selective for aldehydes and other VOCs


Featured

1. Selective trace level monitoring (LOD) with detection limits below 0.1 ppb for CF4, with user-configurable response times from 10 seconds to several minutes.

2. Wide dynamic range, stable operation with no consumables required

3. Low sample volume ( seldom milliliters) built-in gas exchange system with long re-calibration intervals, user-configurable monitoring tasks, and an intuitive user interface.

4. The built-in display displays the results of remote operations digitally and graphically via a local network.