Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and measurement services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We hold the necessary qualifications to serve nationwide, addressing clients' electrical safety and explosion-proof fire prevention issues, and contributing to safe production.
How often should various gas detectors be calibrated and tested?
The recalibration period for instruments is typically one year, which is determined by the submitting unit based on actual usage. During the secondary calibration process, if there are doubts about the instrument test data, or if the main components of the instrument are replaced, or if recalibration is required after repair.
Regular inspection items for environmental gas detection and alarm instruments in workspaces:
1. Alarm sound and light (if applicable)
2. Low response value
3. Indicated Error (for Gas Detectors/Gas Detection Alarms)
4. Instrument Drift (in Portable Instruments)
5. Alarm Error (for Gas Detection Alarm / Gas Alarm)
6. Pump Suction Instrument Gas Sampling Distance and Flow Rate (for Pump Suction Instruments)
7. Power Supply (for battery-powered instruments)
8. Response Time
9. Status Indicator (with wireless communication function)
Gaseous Alarm Meters Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
2. Relative Humidity: ≤85%
3. The working environment should not interfere with the normal operation of the instrument through electromagnetic fields and interfering gases.
Calibration Tools and Instruments
1) Standard Gas Substance
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Unit: Dilution Error of 1%
3) Stopwatch: Graduation not exceeding 0.1 seconds.
4) Flow Control: Flow rate not less than 500 mL/min, flow grade not less than level 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that matches with the calibrated cylinder.
7) Gas Pipeline: Tubing made of materials like polytetrafluoroethylene or stainless steel that do not affect the concentration of the measured gas.
Regulatory Content and Methods for Regular Detection of Flammable Gas Detectors
Annually, we conduct annual inspections on the sensors of commercial and industrial alarm systems, primarily checking for the accuracy, repeatability, drift, and response time of the alarms to ensure they meet standards. Additionally, we verify the suitability of the alarm installation location and whether it is affected by interference from other equipment.
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas alarm instruments, providing regular inspections, metrological measurements, and calibration services. We are qualified to address fire and explosion prevention issues for our clients.