Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, verification, and measurement services for gas alarms, combustible gas detectors, toxic and harmful gas detectors, and other equipment at various locations. With qualifications to serve nationwide, we address various electrical safety and explosion-proof fire prevention issues for clients, contributing to safe production and operation.
Gas Detector Alarm Calibrator Specification Requirements
Regulation 5.5 of the "Regulation for Calibration of Combustible Gas Detectors" (JJG693-2011) stipulates: The calibration cycle for combustible gas detection and alarm instruments: generally, the calibration cycle of the instrument should not exceed one year. If there is any doubt about the instrument's measurement data, or if the instrument has been replaced with major components or repaired, it should be sent for calibration promptly.
Section 3.5 of the "Regulations for the Use of Combustible Gas Detection and Alarm Systems" (SY6503-2000) stipulates: Combustible gas detection and alarm systems in use shall be calibrated regularly as per the regulations.
Occupational Health Services - Technical Regulations (GBZ/T223-2009) Article 7.2.1 stipulates: Calibration of detection and alarm devices shall be performed regularly according to the manufacturer's specified time and procedures.
Section 3.4 of the "Regulations for the Use of Combustible Gas Detectors and Alarms" (SY6503-2000) stipulates: Combustible gas detectors and alarms for installation and use shall have measuring instruments manufactured by recognized institutions with ** explosion-proof performance **.
Which gas detectors require regular calibration and verification?
All types of in-service fixed, handheld, portable gas detectors, gas alarm instruments, toxic and harmful gas detectors/alarm, flammable gas detectors/alarms, gas analyzers, etc.
Gaseous Alarm Detector Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
Relative Humidity: ≤85%
3. The working environment should not interfere with the normal operation of the instrument through electromagnetic fields or interfere with gases.
Inspection Gauges and Instruments
1) Certified Reference Materials for Gases
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Device: Dilution error of 1%.
3) Stopwatch: Graduation not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than level 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that matches with the calibrated cylinder.
7) Gas Pipeline: Tubing made of materials such as polytetrafluoroethylene or stainless steel that do not affect the concentration of the measured gas.
How often should various gas detectors be periodically calibrated and tested?
Instrument recalibration is typically every 1 year, determined by the submitting unit based on actual usage. During the secondary calibration process, if there are doubts about the instrument's test data, or if the main components of the instrument are replaced, or if recalibration is needed after repairs.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is an independent third-party service provider specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic gas detection equipment. We offer regular calibration, measurement, and certification services, as well as qualifications to assist clients in solving fire and explosion prevention issues.