Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the certification, calibration, and metrological services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We are fully licensed, serving the entire nation, and dedicated to solving electrical safety and explosion-proof fire prevention issues for our clients, supporting safe production.
Regular inspection content for gas alarms
Installation spacing of inspection (probe) points
2. Installation Height of the Detector
3. Indicated Error (Concentration)
4. Repeatability (Concentration)
5. Alarm Error (Concentration)
6. Response Time (Seconds)
How often should various gas detectors be periodically calibrated and tested?
The recalibration period for instruments is typically one year, determined by the submitting unit based on actual usage. During the secondary calibration process, if there are doubts about the instrument's test data, or if the main components of the instrument are replaced, or if recalibration is needed after repairs.
Which gas detectors require regular calibration and verification?
Fixed-install, handheld, and portable gas detectors, gas alarms, toxic and harmful gas detectors/alarm systems, flammable gas detectors/alarm systems, gas analyzers, etc.
Gas Detector Calibration: Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃
Relative humidity: ≤85%.
3. The working environment should not interfere with the normal operation of the instrument through electromagnetic fields and interfering gases.
Calibration Tools and Instruments
1) Certified Reference Materials for Gases
The relative expanded uncertainty of the gas standard substance does not exceed 2% (k=2).
2) Standard Gas Dilution Device: Dilution error of 1%.
3) Stopwatch: Graduation not greater than 0.1s.
4) Flow Control: Flow range not less than 500 mL/min, flow grade not less than level 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Relief Valve: A pressure relief valve that is compatible with the calibration cylinder.
7) Gas Pipeline: Tubing made of materials like polytetrafluoroethylene or stainless steel that do not affect the concentration of the tested gas.
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proof, gas alarm, and toxic/hazardous gas alarm instrument calibration, metrological verification, and regular inspections. We hold the necessary qualifications to address fire and explosion prevention issues for our clients.