Shenzhen Zhongnuo Testing Technology Co., Ltd. specializes in the calibration, certification, and measurement services for gas alarms, combustible gas detectors, toxic and harmful gas detectors, and other equipment for various locations. We serve nationwide, addressing clients' electrical safety and explosion-proof fire prevention issues, and contributing to safe production.
The instrument's detection principle primarily includes catalytic combustion, infrared absorption, and thermal conductivity types. Sampling methods include diffusion and aspiration. The instrument is mainly composed of detection elements, amplification circuits, alarm systems, and displays, used for monitoring the concentration of combustible gases in the environment.
Gas Detector Alarm Calibration Standard Requirements
Regulation 5.5 of the "Regulation for the Verification of Flammable Gas Detectors" (JJG693-2011) stipulates: The verification cycle for flammable gas detection and alarm instruments: Generally, the verification cycle for the instrument should not exceed one year. Instruments with suspected measurement data, instruments that have replaced major components, or instruments that have been repaired should be sent for verification promptly.
Section 3.5 of the "Regulations for the Use of Combustible Gas Detection and Alarm Devices" (SY6503-2000) stipulates: Combustible gas detection and alarm devices in use shall be calibrated periodically as per the regulations.
Occupational Health Service - Technical Requirements for the Installation of Toxic Gas Equipment (GBZ/T223-2009) Article 7.2.1 stipulates: Calibration of detection and alarm devices should be carried out regularly according to the manufacturer's specified time and procedures.
Regulation 3.4 of the "Code for the Use of Combustible Gas Detectors and Alarms" (SY6503-2000) stipulates: Combustible gas detectors and alarms for installation and use should have measurement instruments manufactured by an institutionally recognized manufacturer, with explosion-proof performance.
Appearance and structure
The instrument should not have any appearance damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and glossy, with a uniform paint and coating, free of peeling and rusting.
The instrument's connections are reliable, and all dials or buttons should operate and control normally.
Logos and Branding Marks
Instrument name, model, manufacturer's name, date of manufacture, serial number, explosion-proof marking and number, as well as the manufacturing metrological instrument license marking and number for domestic instruments should be clear.
Gas alarm meter calibration conditions and requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
2. Relative Humidity: ≤85%
3. The working environment should not affect the normal operation of the instrument with electromagnetic fields and interfering gases.
Calibration Tools and Instruments
1) Certified Reference Material for Gases
The relative expanded uncertainty of the gas standard material does not exceed 2% (k=2).
2) Standard Gas Dilution Unit: Dilution Error of 1%.
3) Stopwatch: Graduation not greater than 0.1s.
4) Flow Control: Flow rate not less than 500 mL/min, flow grade not lower than Level 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that matches with the calibrated cylinder.
7) Gas Pipeline:气管, made of materials like polytetrafluoroethylene or stainless steel, which do not affect the concentration of the gas being tested.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party service provider specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas detection instruments for regular inspections, metrological measurements, and calibrations. We hold the necessary qualifications to assist customers in addressing fire and explosion prevention issues.