I. Overview
The DWF-10E Constant Current Locatable Large Flow Dust Sampler is a patented instrument independently developed and manufactured by Yancheng Dewang Instrument Equipment Co., Ltd. It is designed to collect total dust and respirable dust in the air, suitable for fixed-point dust sampling of occupational hazards in workplaces, as well as metal dust particles. It is also applicable to routine and emergency monitoring of environmental atmosphere, public health, and occupational health.
The Constant-Current Positionable Large-Flow Dust Sampler is mainly composed of: operating system, intelligent control system, air pump, flow measurement system, and positioning data upload module. It can continuously operate for over 8 hours on a single charge, convenient for sampling tasks. The instrument features real-time display of flow rate, sampling volume, sampling time, temperature, atmospheric pressure, and battery level. It is compact in structure, small in size, lightweight, durable, with strong load-bearing capacity, stable and accurate flow rate, and low noise. It is portable for easy use during sampling.
Section II: Application Fields
The DWF-10E Constant-Current Positionable Large-Flow Dust Sampler is an ideal instrument for short-term or long-term air sampling by environmental protection personnel in disease prevention centers, environmental monitoring stations, colleges and universities, research institutions, and testing organizations.
Standard of Execution:
JJG956-2013规程 for Atmospheric Sampler Calibration
GBZ/T300-2017 Specification for Sampling of Harmful Substances in Workplace Air
GB50325-2020 Specification for Indoor Environmental Pollution Control of Civil Engineering Construction Projects
HJ 194-2017 "Technical Specification for Manual Monitoring of Ambient Air Quality"
III. Technical Specifications
Flow Range: (0-15 L/min) (Flow adjustable)
2. Constant Current Mode: Electronic digital constant current, supporting functions such as data entry, mobile data collection, historical data download, data logic verification and audit, statistical analysis, and visual display.Equipped with a digital DTU positioning module, it allows for real-time upload of sampling data, fault alarms, and display of sampling location's latitude and longitude.Flow is unaffected by voltage fluctuations and air resistance changes.。
3. Flow Error: ≤±2.5%
4. Traffic Stability: ≤3%
5. Traffic Repetition: ≤2%
6. Timing Error: ≤±0.01%
7. Timing Range: 1min to 99h 59min
8. Operating Temperature: -10~50℃
9. Relative Humidity: ≤95%RH
10. Battery Life: Not less than 8 hours
11. Load Capacity: ≥ 30,000 Pa
12. Sampling Mode: ① Cyclic Sampling (99 cycles), with options for cyclic volume or time sampling. Set the sampling time, interval time, and number of cycles.
②Manual sampling; after setting the sampling flow, sampling can be directly initiated, and the timer starts. Manual shutdown available.
③ Timed sampling, set the sampling time, and the sampling stops once the time is reached.
④ Constant-volume sampling (1000L), set the sampling volume.
12. (1) Lock Function: The panel lock feature prevents changes to the settings during the pump's operation.
12 (2) Flow Fault Correction: Fault alarm function display activates when flow variation exceeds ±5% within the specified back pressure range.
After 30 seconds, the error cannot be corrected, and the pump will cease operation.
12 (3) The extra-large LCD screen displays: flow rate, sampling time, sampling volume, back pressure, battery estimated sampling time, date, time, operation mode, and remaining battery power.
12(4) Humidity Range: Operating humidity: 5-95%, non-condensing; Storage humidity: 5-98%, non-condensing
12(5) You can view the relevant working data and actual conditions of the sampler in real-time on our provided cloud platform.
13. Host unit size: 130*44*95mm (L*W*H)
14. Operating Noise: <65 dB
15. Instrument Weight: 1.5 kg
16. Power: DC7.4V, 220VAC±10%
17. Volume Calculation: Real-time Display
18. Data Storage & Download: 1000 sets of real-time queryable sample data with USB port for download
19. Explosion-proof Certificate No.: CNV20.0235 Explosion-proof Mark: Ex ib IIB T4 Gb
1. Main Configuration:
Sampler main unit with battery, dedicated charger, operation manual, carrying case, tripod stand, exhalation dust sampling head, all dust sampling head, exhalation dust clip, all dust clip, impact film, impact film box, silicone oil, tweezers, and other specialized tools, certificate of conformity, warranty card.