Shenzhen ZhongNuo Inspection & Technology Co., Ltd. specializes in the calibration, verification, and measurement services for gas alarms, combustible gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We hold the necessary qualifications to serve nationwide, providing solutions to electrical safety and explosion-proof fire prevention issues for our clients, and contributing to safe production.
Combustible Gas Detector Alarm Regular Inspection Content and Methods
Annually, we conduct annual inspections on the sensors of commercial and industrial alarm systems, primarily checking for the accuracy, repeatability, drift, and response time of the alarms to ensure they meet standards. Additionally, we verify the appropriateness of the alarm installation locations and whether they are subject to interference from other equipment.
Regular Inspection Items for Workplace Environmental Gas Detection and Alarm Equipment
Alarm sound and light (if applicable)
2. Low response value
3. Indicating Error (for Gas Detectors/Gas Detector Alarms)
4. Instrument Drift (in Portable Instruments)
5. Alarm Error (for Gas Detector/ Gas Alarm)
6. Pump-suction instrument gas collection distance and flow rate (for pump-suction instruments)
7. Power Supply (for battery-powered instruments)
8. Response Time
9. Status Indicator (with wireless communication capability)
Appearance and structure
The instrument should not have any surface damage that affects its normal operation. The surface of newly manufactured instruments should be smooth and clean, with a uniform paint and coating, and no signs of peeling or rust.
The instrument's connections are reliable; all dials or buttons should operate and control normally.
Logos and symbols
Instrument name, model, manufacturer's name, date of manufacture, serial number, explosion-proof marking and number, as well as the manufacturing measurement instrument license marking and number for domestic instruments should be clear.
The detection principle of the instruments mainly includes catalytic combustion, infrared absorption, and thermal conductivity types. The sampling methods are diffusion and aspiration. The instruments are primarily composed of detection elements, amplification circuits, alarm systems, and displays, used for monitoring the concentration of combustible gases in the environment.
Shenzhen Zhongnuo Inspection Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic and harmful gas detectors for regular inspections, metrological measurements, and calibrations. We provide certification services to address fire and explosion prevention issues for our clients.