Shenzhen Zhongnuo Testing Technology Co., Ltd. specializes in the calibration, verification, and metrological services for equipment such as gas alarms, flammable gas detectors, and toxic gas analyzers for various locations. We hold qualifications to serve nationwide, providing solutions for electrical safety and explosion-proof fire prevention issues, contributing to safe production.
The calibration cycle for flammable gas detectors is typically once a year.
According to the standard JJG 693-2011 "Flammable Gas Detectors," the calibration cycle for flammable gas alarms generally does not exceed 1 year. This means that under normal circumstances, a comprehensive calibration and test of the flammable gas alarm is required annually to ensure its performance meets the relevant requirements.
GB 12358-2024 "General Technical Requirements for Environmental Gas Detection and Alarm Instruments in Workplaces" is an important standard.
Intended to standardize the performance, functions, and testing methods of environmental gas detection alarm instruments used in work sites.
The standard was issued in 1990 and has undergone several revisions.
The new version is GB 12358—2024, which will be officially implemented on June 1, 2025.
This standard applies to the production and use of gas detection and alarm instruments for flammable gases, toxic and harmful gases, and oxygen in workplace environments.
Routine Inspection Items for Environmental Gas Detection and Alarm Equipment in Workplaces:
Alarm sound and light (if applicable)
2. Low response value
3. Indication Error (for Gas Detectors/Gas Detector Alarms)
4. Instrument Drift (in Portable Instruments)
5. Alarm Error (for Gas Detection Alarm/ Gas Alarm)
6. Pump-actuated instrument gas collection distance and flow rate (for pump-actuated instruments)
7. Power Supply (for battery-powered instruments)
8. Response Time
9. Status Indicator (with wireless communication capability)
Regular inspection content for gas alarms
Installation spacing of inspection (testing) points
2. Installation Height of the Detector
3. Indicated Error (Concentration)
4. Repeatability (Concentration)
5. Alarm Error (Concentration)
6. Response Time (Seconds)
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarm systems, and toxic and harmful gas detection instruments. We offer regular inspections, metrological services, and calibration. Our资质 enables us to provide solutions for fire and explosion prevention issues to our clients.