Shenzhen ZhongNuo Testing Technology Co., Ltd. specializes in the calibration, certification, and measurement services for gas alarms, combustible gas detectors, toxic and harmful gas detectors, and other equipment, serving various locations across China. We address electrical safety and explosion-proof fire prevention issues for clients, contributing to safe production.
Gas Alarm Meters Calibration Conditions and Requirements
Environmental Conditions
Ambient Temperature: 0~40℃.
2. Relative Humidity: ≤85%
3. The working environment should not affect the normal operation of the instrument with electromagnetic fields and interfering gases.
Inspection Gauges and Instruments
1) Standard Gas Substance
The relative expanded uncertainty of the gas standard substance does not exceed 2% (k=2).
2) Standard Gas Dilution Device: Dilution error of 1%.
3) Stopwatch: Graduation not greater than 0.1s.
4) Flow Control: Flow rate not less than 500 mL/min, flow grade not less than 4.
5) Zero Point Gas: Purified compressed air.
6) Pressure Reducing Valve: A pressure reducing valve that is compatible with the calibrated cylinder.
7) Gas Pipeline: Use gas piping made of materials such as polytetrafluoroethylene or stainless steel that do not affect the concentration of the measured gas.
GB 12358-2024 "General Technical Requirements for Environmental Gas Detection and Alarm Instruments in Workplace" is an important standard.
Intended to standardize the performance, functions, and testing methods of environmental gas detection and alarm instruments used in work places.
The standard was released in 1990 and has undergone several revisions.
The new version is GB 12358—2024, which will be officially implemented on June 1, 2025.
This standard applies to the production and use of gas detection and alarm instruments for flammable gases, toxic and harmful gases, and oxygen in workplaces.
Power-on inspection
Upon powering on the instrument, it should operate normally with clear and complete display sections. Check the alarm function and alarm trigger values.
Alarm function and alarm action value verification
The instrument's sound and light alarm should be normal. Insulation resistance
Insulation resistance for instruments using alternating current power should be no less than 20MΩ.
What factors can affect the calibration cycle of a flammable gas detector alarm?
Manufacturer Requirements: Different manufacturers may have varying inspection cycle requirements for their flammable gas detectors. Therefore, when purchasing equipment, carefully review the product manual to understand the specific requirements of the manufacturer.
Equipment Operating Environment: The operating environment of the alarm can directly impact its inspection cycle. If the equipment is exposed to high concentrations of harmful gases, high or low humidity, or high or low temperatures for extended periods, its inspection cycle may be shortened accordingly.
Equipment Maintenance Status: The condition of equipment maintenance is also a significant factor affecting the inspection cycle. If the equipment is frequently facing issues like battery depletion and sensor aging, the inspection cycle will accordingly be shortened.
Shenzhen Zhongnuo Testing Technology Co., Ltd. is a third-party institution specializing in electrical safety, electrical explosion-proofing, gas alarm systems, and toxic and harmful gas detection instruments for regular inspections, metrological measurements, and calibration services. We are qualified to address fire and explosion prevention issues for our clients.