Shenzhen Zhongnuo Testing Technology Co., Ltd. specializes in the calibration, certification, and measurement services for gas alarms, flammable gas detectors, toxic and harmful gas detectors, and other equipment in various locations. We serve nationwide, addressing customers' electrical safety and explosion-proof fire prevention issues, and supporting safe production.
Measurement Equipment Control
Instrument control includes calibration, follow-up calibration, and in-service inspection.
Regular inspection items for environmental gas detection and alarm instruments in the workplace:
1. Alarm sound and light (if applicable)
2. Low response value
3. Indication Error (for Gas Detectors/ Gas Detector Alarms)
4. Instrument Drift (in Portable Instruments)
5. Alarm Error (for Gas Detection Alarm/ Gas Alarm)
6. Pump-suction instrument gas collection distance and flow rate (for pump-suction instruments)
7. Power Supply (for battery-powered instruments)
8. Response Time
9. Status Indicator (with Wireless Communication Function)
How often is it required to regularly calibrate and check various gas detectors?
Instrument recalibration is typically every 1 year, determined by the sending unit based on actual usage. During the secondary calibration process, if there are doubts about the instrument's test data, or if the main components of the instrument are replaced, or if recalibration is required after repairs.
Flammable and toxic/hazardous gas alarm detectors verification and metrological calibration according to standards:
Standard GB/T50493-2019 "Design Standard for Detection and Alarm Systems of Combustible Gases and Toxic Gases in Petroleum and Chemical Industry"
GB12358-2006 General Technical Requirements for Environmental Gas Detection and Alarm Instruments in Workplace
JJG693-2011 "Regulation for Calibration of Combustible Gas Detectors"
Shenzhen ZhongNuo Testing Technology Co., Ltd. is a third-party service provider specializing in electrical safety, electrical explosion-proofing, gas alarms, and toxic gas detection instruments. We offer regular calibration, measurement, and certification services, and are qualified to address fire and explosion prevention issues for our clients.