PET-1/2 Oil Drive System Filtration System Evaluation Device
Introduction:
This instrument is primarily used for measuring the filtration flow rate and total filtration volume of a liquid subjected to a certain pressure as it passes through a specific filter membrane over time., Capillary shear tests can also be conducted. Widely applicable in medical, chemical, and oilfield industries. For instance, in the oilfield sector, it can be used for evaluating multiple parameters of polymers, including filtration factor and continuous filtration ratio, particularly suitable for measuring the filtration factor in the enterprise standard -- Q/SY DQ0712-2000 "Technical Requirements for Partially Hydrolyzed Polyacrylamide Used in Oil Displacement." The filtration factor is an empirical constant used to measure the uniformity of polymer solutions.
Primary Applications: Polymer filtration factor testing, water-insoluble matter experiments, capillary shear tests, packed sand tube shear tests, wastewater filtration experiments, core shear evaluations, and other related experiments.
Product Description: Feature Introduction
Four measurement units, capable of conducting four independent filter flow tests simultaneously.
The container components feature a hatch-type sealing structure, which can be easily opened and sealed by pushing and pulling the handle, making the installation, removal, and filling of liquids in the container components both convenient and effortless.
Filter membrane holder components are connected with quick couplings for easy and quick filter membrane replacement.
Utilize a weight flow meter to measure filtrate flow and total volume, enhancing measurement accuracy.
Micro, silent air compressors are used for the gas source.
Independent design rock core holder, easy to install and operate.
Can independently achieve sand filling tube simulation experiments, etc.
Can be used for custom high borosilicate sand chips and other experiments.
Capillary shear test data;
The main control unit of the measuring device employs an AVR (microprocessor). The dedicated control system allows for independent control and measurement of the four measuring units, featuring constant filtering pressure control and protection for open sample containers.
The measuring device's main unit communicates with the PC via USB interface, operates dedicated control and measurement software under WINDOWS, and can separately control and collect data from four measurement units.
Real-time display of time-flow and time-total curves. Capable of printing time-flow and time-total curves, with the filter factor marked.
Instrument measurement parameters
Gas Source:
Types: Compressed Air, utilizing mini, silent air compressors; (Instrument built-in air source)
Gas supply rate: Not less than 12 L/min
Pressure Provided: Not less than 0.6 MPa
2. Airway Pressure Control
Constant Range: 0MPa-0.3MPa; Set Increment: 0.01MPa
Pressure Control: ±1%
Flow rate at constant pressure: Not less than 10ml/sec
Pressure Units: Mpa, PSI, Kg/cm²
3. Container Components:
Container Volume: 500ml
Material: Acrylic
Pressure Resistance of Container: Not less than 0.5 MPa
Clamping Material: 316L Stainless Steel; Filter Film Clamp: Φ47mm
Clamping Connector Type: The filter membrane clamping assembly uses quick connectors for easy and quick filter membrane replacement; (includes technical patent)
4. Software Features:
Flow Rate Measurement: Not less than 10ml/sec
Volume detection sensitivity: 0.5ml
Measurement Time: Sampling Interval 0.1-10 sec (0.1 sec increment setting)
Total Measurement Time: Not less than 100 hours, timing accuracy: 0.1s
Measurement Units: Four, capable of conducting four independent filter flow tests simultaneously
Communication Function: The main unit of the measuring device communicates with the PC via USB interface, operates dedicated control and measurement software under WINDOWS, and can individually control and collect data from the four measurement units.
Data Function: Real-time display of time-flow and time-total curves. Ability to print time-flow and time-total curves, with marked filter factors.
5. Weight-Flow Meter Function:
Utilize a weight flow meter to measure filtrate flow and total volume
Weight Flow Meter: Automatically records flow time and calculates the filter factor value.
No computer connection required; operates independently and displays filter factor values.
6. Customization Features:
Sand filling tube shear test data function; (Customized) Capillary size: 200*1 (Customized)
7. Power Voltage:
AC220V 50Hz; Power: ≤500W; Starting Current of Air Compressor: < 10A
8. Dimensions:
650x640X900(mm)
9. Net Weight of Main Unit:
52 Kg (excluding flow meter and container components)