Features and Technical Specifications
5-inch resistive touch screen monitor, all-Chinese operating interface, user-friendly program design.
● From 1 to 7 points calibration mode, automatically calculates slope, intercept, and correlation coefficient, with high measurement accuracy.
● Large capacity data storage with power interruption protection design ensures that instruments are not damaged and data records are not lost.
● High anti-interference capability, suitable for industrial environments.
● Widely applicable for monitoring surface water and pollution sources.
● Features a unique semiconductor cold light source emitter with a lifespan of up to tens of thousands of hours.
● Data can be exported via a USB flash drive, with simple operation and convenient portability.
● Low reagent consumption, low operation cost, and strong anti-interference ability.
System Parameters
COD:
1 Method of Measurement: Rapid Catalytic Method (Chromium Method)
2 Measuring Range: 5-1000 mg/L (dilute and measure when >1000 mg/L)
3 Measurement Error: 5-100 mg/L, error ≤ ±10 mg/L; 100 mg/L-1000 mg/L, relative error ≤ ±5%.
4 Dissolution Temperature: 165 ± 1℃; Dissolution Time: 10 minutes;
5 Chlorine Interference: [CL-] <1000mg/L;
6 Curve Parameters: Settable for 60 measurement curve parameters;
7 Record Storage: Stores over 50,000 measurement results.
Data Export: Exportable to USB flash drive.
Ammonia Nitrogen:
1 Method of Measurement: Nessler's colorimetry
2 Measurement Range: (0 ~ 10 mg/L)
3 Measurement Error: ≤±5%
4 Curve Parameters: Up to 60 measurable curve parameters can be set.
Calibration: 1-7 point calibration mode, automatic curve value calibration.
6 Curve Parameters: Up to 60 measurement curve parameters can be set.
7 Record Storage: Stores over 50,000 measurement results.
8 Data Export: Exportable to USB flash drive.
Ambient Temperature: (5~40)℃; Relative Humidity: <85% (no condensation)
Total Phosphorus:
① Determination Method: Ammonium molybdate spectrophotometric method
② 2 Measuring Range: 0.01 mg/L to 0.6 mg/L
③ Measurement Error: ≤±5%
④ 4 Decomposition Temperature: 120±1℃; Decomposition Time: 30 minutes;
⑤ 5 Curve Parameters: Up to 60 measurement curve parameters can be set.
⑥6 Record Storage: Stores over 50,000 measurement results.
⑦ Data Export: Exportable to USB flash drive.
⑧8 Ambient Temperature: (5~40)℃; Relative Humidity: < 85% (No Condensation)
Total Nitrogen:
1 Method of measurement:光度比色法
2 Measuring Range: ≤ 120mg/L; Measurement Error: ≤ ±5%
3 Decomposition Temperature: 125±1℃; Decomposition Time: 30 minutes
4 Curve Parameters: Up to 60 measurable curve parameters can be set.
5 Record Storage: Stores over 50,000 measurement results
6 Data Export: Exportable to USB flash drive;
7 Ambient Temperature: (5~40)℃; Relative Humidity: <85% (no condensation).