Principle and Introduction of pH Electrode
The principle of pH electrode measurement is electrochemical method and the principle of primary battery. primary cell
It is a system that converts chemical energy into electrical energy. This electricity
The voltage of a pool is called electromotive force (EMF). This electromotive force (EMF) is generated by
Composed of two and a half batteries. One of the half cells is called the measuring cell, and its
The potential is related to specific ion activity; The other half cell is the reference half cell
A pool, commonly referred to as a reference electrode, is typically connected to the measuring solution and
And it is connected to the measuring instrument. Electricity generated through the reaction of the primary battery inside the electrode
The potential difference is transmitted to the pH controller, and the corresponding algorithm sends the displayed pH value.
1
Precautions for using pH 2 electrodes
Precautions for using pH 2 electrodes
(1) The electrode leads are special shielded wires, and customers are prohibited from splicing leads without authorization
We are not responsible for any consequences caused by the line.
(2) Long term immersion in distilled water or protein solutions should be avoided, and
Prevent contact with organic silicone grease.
(3) The glass film of electrodes that have been used for a long time may become transparent
If there is sediment attached, it can be washed with a dilute solution and rinsed with water.
(4) When you perform maintenance and upkeep on the electrodes, you still cannot calibrate them
The program and normal testing indicate that the electrode is no longer able to recover its response. Please replace the battery
Extreme.
(5) The pH electrode wire is not waterproof and should be avoided from coming into contact with water as much as possible.
(6) The shelf life of the electrode is one year, and it should be used as soon as possible after purchase,
Not suitable for storage for too long.