German Leybold Resistance Vacuum Gauges/Vacuum Gauge Tubes/Vacuum Instruments DISPLAY one
4.1.2 Environmental Requirements
Temperature: Storage: -20...+60℃
Operation: +5...+50℃ (at sea level)
+5…+30°C (at an altitude of 2,000 meters)
Relative Humidity: 80% (within 30℃)
Reduced to 50% (above 40℃)
Indoor (Altitude up to 2,000 meters)
Pollution Level: II
Protection Class: IP40
4.1.3 Standard
■ Compliant with Low Voltage Directive 2009/95/EC
■ Compliant with Electromagnetic Compatibility Directive 2004/108/EC
International Standards / National Standards and Technical Specifications
■ DIN EN 61010-1
Safety Requirements for Electrical Equipment Used in Measurement, Control, and Laboratories
■ DIN EN 61000-6-2
Electromagnetic Compatibility (EMC)
■ DIN EN 61000-6-3
Electromagnetic Compatibility (EMC)
4.2 Power Connection
Voltage: 100...240 VAC
Frequency: 50/60 Hz
Circuit Breaker: 2x T1.6A
Power Consumption: 5VA
Overvoltage Category: II
Protection Category: 1
Connection: Connector for non-heating equipment IEC 320 C14
4.3 Measurement Channel
Quantity: 1
Connect: RJ45 (FCC68)
12
Regulatory Connectors: THERMOVAC TTR90, TTR90S
TTR91,TTR91S
TTR96S
TTR211S
TTR216S
TTR100,TTR100S2
PENNINGVAC PTR90
4.3.1 Regulatory Power Supply
Voltage: +24VDC ±5%
Current: 100mA
Circuit Breaker: 200mA Self-Protection (Power supply complies with low-voltage electrical equipment grounding protection requirements EN61010)
4.3.2 Measurement Parameters
Range: Determined by the connected gauge (1.5E3…5E-9 mbar)
Measurement Rate: 50 s^-1
Display Rate: 4s^-1
Measurement Units: Millibars, Pascals, Tορs
4.3.3 Switches / Relay Outputs
Contact Type: Transformer Contact, Floating
Load (resistive): Switch Current: 1A
Switching Voltage: 30VAC/60VDC
Service Life: Mechanical: 5.10 7 Switching Cycle
Electrical: 10-4 switch cycle under load
Connection: Wiring Terminal
4.3.4 Analog Output
Quantity: 1 per measurement channel
Voltage Range: 0...10VDC
Display Value Deviation: ±0.1%