详情描述

FT-8600 Series Automatic Powder Resistivity and Compaction Density Meter - Powder Conductivity Measurement


OneFeature Description:

Offering new energy battery material powder compaction density and resistivityMeasurement of conductivity and foil resistance; AutomaticServo Power SystemPowderReal-time graph of the relationship between compaction resistance and density, and pressure versus compaction density graphAutomatically analyzes the deformation after constant pressure and pressure release of the sample, as well as the resistivity and conductivity profiles during the elastic deformation process, and measures the height of the sample.0.001mm; Two operating modes: touch screen and PC software; Integrated control system, reporting features; Automatic operation; The instrument has self-protection and abnormal alarm functions.

TwoReference Standard:

FT-8600 Series Automatic Powder Resistivity and Compaction Density Meter - Powder Conductivity Measurement

1. Density Measurement Reference: GBT 24533-2009 Graphite Anode Materials for Lithium-Ion Batteries; GB/T 24533-2019 Graphite Anode Materials for Lithium-Ion Batteries

AppendixTesting Method for Powder Compaction Density

2. Four-probe tester in accordance with GB/T 1551, GB/T 1552-1995, and ASTM F84 American A.S.T.M. standard.

3. The four-probe method for powder testing is in accordance with the requirements for the four-probe method in the arbitration method, as specified in the GBT 30835-2014 "Lithium-ion Battery Carbon-Composite Phosphate Iron Lithium Cathode Material" regarding the determination method of powder electrical conductivity.

ThreeApplication Scope:

Measurement of the resistivity for the lithium-ion battery cathode material, lithium iron phosphate, in carbon composite form.

2. Analysis and testing of semiconductor powder materials requiring the four-probe method measurement.

3. For measuring the resistivity of powdered materials like graphite.

4. Film Resistor Measurement

5Used for corporate quality inspection; R&D departments, colleges and universities; research institutions; and quality testing agencies.

FourPrinciple of Operation:

During the compression process by external forces, as the powder moves and deforms, the voids are filled, the contact area between particles increases, causing an attraction between atoms and strengthening the mechanical wedging between particles, thereby forming a compact with density and strength.Real-time measurement of resistance, resistivity, and conductivity changes during the forming of the blank until it stabilizes.

FiveParameters and Specifications







FT-8600 Series Automatic Powder Resistivity and Compaction Density Meter - Powder Conductivity Measurement

Test Item/ModelFT-8600AFT-8600B

Resistor/Sheet Resistance Range10^-5~2×10^5Ω/□10^-4~1×10^7Ω/

2. Resistance range10^-6~2×10^6Ω-cm10^-5~2×10^8Ω-cm

3. Conductivity Range5×10^-6~10^6ms/cm5×10^-8~10^5ms/cm

4. Test Current Range0.1μA ,1μA,10μA,100µA,1mA,10mA,100 mA10mA ---200pA

5. Current Accuracy±0.1%±2%

6 Main Pressure Selection2 tons, 3 tons, 5 tons

7 Sample Height Measurement0.001mm—25.000mm; Automatic

8-bit displacement0-50mm

Control SystemServo Power Control System

Display Method7-inch touchscreen + PC software

11. Compaction Molds(National Standard) Inner Diameter: 13mm; Height: 25mm; Precision: 0.01mm

Resistivity Measurement MoldingInner diameter: 16mm; Height: 10mm (recommended height after sample compression ≤ 4mm)

Polarized Film Resistance ElectrodeStandard Configuration: Electrode Diameter 16mm; other specifications available for customization.

12. Molding Release MethodAutomatic

13. Protective DeviceAutomatic diagnostics, overload protection, and emergency stop control

14. Temperature and Humidity RangeRoom temperature -50°C; Humidity: 20%-98%

15. Time Settings0-99.9S

16. Density Value99.99999 g/cm³ - Different samples have varying densities; Tolerance: 1%

17. Deformation VariableMeasurable deformation of elastic materials

18. Resistance Testing MethodsStandard four-probe method; four-point method (optional with the corresponding measuring electrode)

19. Resistance Repeatability Error≤0.3% (Standard Resistance) Repeatability error ≤2%

(Subject to the characteristics of sample variability and uniformity, etc.)

20. Displacement Speed<15mm/min;0.1mm/min

21. Environmental RequirementsInstrument operating environment requirements: Operating temperature: -10℃ to 40℃, humidity: ≤65%RH

22. Power Supply380v

23. Balance Scale (Select)0.00001g, Range: 120g; 0.01%