详情描述
1. Key Features:
Measure the roundness and coaxiality of measurable circular workpieces.同心度: concentricityTactile count, verticality, parallelism, planarity, shaft弯曲度、Surface waviness, spectral analysis, wave height analysis, etc.
Roundness Filter Level:1-500、1-250、1-150、1-50、1-15、3-16、3-17、15-500、15-250、
Filter Type: GaussianISO Standard, 2RC
2. Key Technical Parameters:
Measurement Range | Measurable Diameter | Ф300mm |
Measurable inner diameter | Ф5mm | |
Measurable height | 400mm | |
Spindle Accuracy | Radial Error | ±(0.025+5H/10000)μm |
Axial Error | ±(0.025+6X/10000)μm | |
Workbench | Top Surface Diameter | φ150mm |
Revolving Diameter | φ300mm | |
Carrying Capacity | 25Kg | |
Adjustment Range | Adjust eccentricity±2mm; Leveling ±1° | |
Rotational Speed | 0-12 RPM | |
Horizontal arm | Horizontal Travel Distance | 150mm |
Mobile speed | 0.5-6 mm/s or manual | |
Sensor | Range | 500μm (radius difference) |
Resolution | 0.005μm | |
Measuring needle shape | Φ2MM Gem Ball Probe | |
Force Gauge | 1-12g | |
Data Collection | Imported Grating14400:00/week | |
Magnification Ratio | At will to200,000x | |
3.Principle of Measurement:
Using the radius measurement method, the workpiece rotates. The rotating axis system employs a high-precision air-bearing spindle as the measurement reference; the electrical part consists of advanced computers and imported precision circular grating sensors, as well as precision inductive displacement sensors; the imported precision circular grating sensors and precision inductive displacement sensors measure angles and radial displacement, ensuring the accuracy of the angular displacement and radial values of the measured workpiece; the measurement software is based on the Chinese versionDTP system measurement software for Windows operating platform, which handles data collection, processing, and printing output.
4.Primary Configuration:
Mechanical parts | Mainframe instrument rack | |
Marble worktop | ||
Air Float Main Shaft System | ||
Transverse guide rail moving device | ||
Air Filtration System | ||
Adjust and level Workbench | Counter DiameterФ180mm, Countertop Height: 50mm | |
Adjustment Range: Offset Adjustment±2mm; Leveling ±1° | ||
Adjustable distance in X-direction: 0-4mm; Slope ratio: 150:3 | ||
Adjustable distance in Y-direction: 0-4mm; slope ratio: 150:3 | ||
Precision Three-Jaw Chuck | External size clamping range:ф0.8 mm ~~ф63mm | |
Internal dimension clamping range:ф23 mm ~~~ф58 mm | ||
Internal Diameter:26 mm | ||
Overall height including flange:42mm | ||
Collet Weight:1.2KG | ||
Electrical appliances section | Electrical Control Systems | |
Import Rotating Circular Encoder | ||
Inductive sensor | ||
Industrial Control Computer | Industrial Control Motherboard NORCO970 | |
Industrial Control Board NORCO PBP-14P4 | ||
CPU E5500 | ||
Kingston 4G Series 3rd Generation | ||
Hard DriveSATA 1TB | ||
19Polarized LCD | ||
Based onwin 10System DownDTP measurement software | ||
HP Color Printer | ||
Standard Components | Standard Glass Hemisphere | |
Elliptical standard parts | ||